SOP for Calibration of Head Space Gas Chromatograph


Standard operating procedure to verify the performance of Head Space Gas Chromatography.

1.0  OBJECTIVE 

       To calibrate the equipment for reliable and accurate results.

2.0  SCOPE

       This procedure is applicable to calibrate Head Space Gas Chromatograph installed in Quality Control Laboratory.

3.0  Responsibility

3.1  Doing      : Technical Assistant
3.2  Checking : Executive/Manager

4.0  Accountability 

       Head of the Department

5.0  Procedure

5.1  (I) BY MANUAL INJECTION :

       Frequency : Every 3 month.
5.1.1  Install the DB-5 capillary column.
5.1.2  Set all parameters as per Annexure – I.
5.1.3  Stabilized the column for stable base line.
5.1.4  Inject 1 µl of each standard solution in duplicate.
5.1.5  Calculate the ratio of Area of methanol to internal standard.
5.1.6  Record the results in Annexure – I.
5.1.7  Calculate the linearity and corelation coefficient.

5.2  (II) FOR HEAD SPACE INJECTION :

       Frequency : Every 6 months.
5.2.1  Install the DB-5 capillary column.
5.2.2  Set all parameter as per Annexure – II
5.2.3  Prepare the standard solution as mentioned in Annexure  – II
5.3  Check the performance for
(1)  Reproducibility test
(2)  Linearity test.

5.4  REProducibility test

5.4.1  Take 5 ml of standard preparation in 5 Head Space vial.
5.4.2  Add 1 gm of sodium sulphate (Anhydrous) in each.
5.4.3  Inject 1 ml gaseous form of standard from each of 5 vials.
5.4.4  Record the result in Annexure  – II
5.4.5  Calculate the standard deviation and related standard deviation for respective separated solvent peak.

Related: Calibration of Gas Chromatography (GC)

5.5  LIinearity test 

5.5.1  Take 2 ml, 5 ml and 10 ml of standard solution in Head Space Vial.
5.5.2  Add 1gm sodium sulphate (Anhydrous) make 10 ml volume by addition of required volume in respective head space vial.
5.5.3  Inject 1 ml Gaseous form of each standard.
5.5.4  Record the result in Annexure – II
5.5.5  Calculate the regression coefficient of respective separated solvent peak.
5.6  After completion of Analysis, cool down the G.C.
5.7  After cooling, switch ‘OFF’ the instrument.
5.8  Switch ‘OFF’ the mains.

6.0  ABBREVIATIONS

6.1  G.C.  = Gas chromatograph
6.2  ml     = millimeter


ANNEXURE-I
QUALITY CONTROL LABORATORY

CALIBRATION  CHECK REPORT OF
GAS CHROMATOGRAPH
CALIBRATION DATE

DATE OF LAST CALIBRATION DONE
NEXT DUE FOR
CALIBRATION




INSTRUMENT DETAILS
INSTRUMENT NAME

INSTRUMENT MAKE

INSTRUMENT
IDENTIFICATION NO.





OBSERVATION

(I)  FOR  MANUAL INJECTION
G.C. Parameters:
Column                    : DB-5 Capillary (0.32mm X 30meter X 0.25micron)                   
Range                      : 2
Oven Temp.            : 300C
Split Ratio               : 1:20
Detector Temp        : 1800C
Nitrogen                  : 100 kPa
Inj. Temp.               : 1400C
Carrier                    : 25Kpa (Helium)
Inj. Volume             : 1.0μl
Hydrogen                : 55Kpa
Air                          : 85 Kpa
Preparation of internal Standard: 25 ml Ethanol è 1000 ml with water
Test preparation : 1 ml, 2 ml, 3 ml, 4 ml, 5 ml Methanol 100 ml with Internal Std.


Methanol (% v/v)

 Ratio of Methanol/Ethanol
 Mean Ratio
1.0
(1)

(2)
2.0
(1)

(2)
3.0
(1)

(2)
4.0
(1)

(2)
5.0
(1)

(2)              
               Correlation coefficient :___________ (Not less than 0.9900)

                       PERFOMED BY

                                       CHECKED BY

NAME
SIGN.
DATE
NAME
SIGN.
DATE