SOP for Linearity Check of High Performance Liquid Chromatograph (HPLC)


Standard operating procedure to check performance of the high performance liquid chromatograph (HPLC) for accurate results.

1.0  Objective

       To check performance of the equipment for reliable and accurate results.                                         

2.0  SCOPE

       This procedure is applicable to check performance of the high performance liquid chromatograph .

3.0  Responsibility  

3.1  Doing      : Technical Assistant/Executive
3.2  Checking : Executive /Manager

4.0  Accountability   

       Head of the Department

5.0  Procedure

5.1  Operate the instrument as per respective standard operating procedure of HPLC. 
5.4  Instrument  Performance ( Linearity of Results)  
       Frequcny :  Two month.
5.4.1  Prepare the mobile phase consisting of 70% Methanol : 30 % water  .
5.4.1  Stock solution preparation : Prepare a mixture  of  0.2 % Benzene and 0.2 % Toluene in methanol.
5.4.2  Make dilution in following manner.
5.4.1  Take 5 ml, 10 ml, 15 ml, 20 ml of stock solution respectively in 25 ml of volumetric solution and make up with methanol.  
5.4.3  Set flow rate at 1.0 ml/min.
5.4.4  Allow the system to be saturated with mobile phase for at least 15 minutes before injecting the different solutions.
5.4.5  Record the area and retention time of both benzene and toluene as per Annexure -I .
5.4.6  Calculate the Linearity for both Benzene and Toluene respectively .

6.0  ABBREVIATION

6.1  SOP - Standard Operating Procedure

Related: Relative Response Factor Calculation in HPLC 


                                                                   ANNEXURE-I
                                                                 Q. C. Department

LINEARITY CHECK REPORT OF HIGH PERFORMANCE LIQUID CHROMATOGRAPH

PERFORMANCE DATE

DATE OF LAST PERFORMANCE
DONE
NEXT DUE FOR
PERFORMANCE




INSTRUMENT DETAILS

 INSTRUMENT NAME

INSTRUMENT MAKE

INSTRUMENT
IDENTIFICATION NO.




OBSERVATION


1) Instrument performance ( Linearity of results)
Stock solution preparation :
Mixture of 0.2 % Benzene and 0.2 % Toluene in Methanol.
Solution 1) : 5.0 ml of Stock solution                     25 ml with methanol.
Solution 2) : 10.0 ml of Stock solution                   25 ml with methanol.
Solution 3) : 15.0 ml of Stock solution                   25 ml with methanol.
Solution 4) : 20.0 ml of Stock solution                   25 ml with methanol.
Solution 5) : Direct stock solution.
HPLC Parameter :
Column         : ODS (25 cm X 4.6 mm, 5 ยต)
Mobile phase : Methanol : Water
                            70     :   30
Flow rate        :  1.0ml/min.
Wave length    :  At 254 nm
Solution
No.
Conc.
Inj.No.
Bin No.
Area of Benzene
Area of Toluene
1
0.04%
1





2



2
0.08%
1





2



3
0.12%
1





2



4
0.16%
1





2



5
0.20%
1





2



Linearity : (Limit NLT 99.95%)