SOP for Environmental Monitoring by Non Viable Air Particle Count


Standard operating procedure to measure non viable count of air in solid dosage form area at rest condition in class 100 and 10000.

1.0  Objective

       To  ensure the control over non viable count in solid dosage form area at rest condition.

2.0  SCOPE

       To measure non viable count of air in solid dosage form area at rest condition. 

3.0  Responsibility

3.1  Doing      :  Technical Assistant./Executive
3.2  Checking : Executive/Manager    

4.0  Accountability

       Head  of  Department    

5.0  Procedure          

5.1  Frequency of the test

Cover all operational area (“O” area) of solid dosage form  once in a three month. ( 3 Months + 7 days )
5.2  Location chart for non viable particle count of air in solid dosage area is attached .
5.3  Instrument Used:- Laser particle counter
5.4  Take air sample from a  location near the concerned  operation at its  working height. Repeat  and take two (2) such reading .
5.5  Sample the air at the rate of 1 ft3 for 1 min.
5.6  Operate the instrument as per S.O.P.
5.7  Record the results in the Annexure-I.

5.8  Acceptance criteria /ft3

Limits
Class
Air particle count
Alert level
Action level
100

> 0.5µ
76 – 100
>100
> 5.0µ
Nil
Nil
100,000
> 0.5µ
80,000 – 100,000
>100,000
> 5.0µ
400 - 560
>560

6.0  Abbreviations

6.1  µm = Micro meter
6.2  ft3 = Feet cube


ANNEXURE – I
                                                       MICROBIOLOGY DEPT                                                      

NON VIABLE PARTICLE COUNT OF  SOLID DOSAGE AREA

                                                                                                            S.O.P.NO.-_________________


Test :- Non  Viable Air Particle Count
Location                           :-
Sampling Point   No.         :-
 Date of sampling  :-
Instrument Id.No :-
 Date of Report      :-
Sampled area       :- 1 ft3
 Sample run time   :- 1 minute

RESULT :-                                                                         

                     SR.No
No. of readings
(Particle / ft 3)
0.5 µ
5.0 µ
1
First reading


2
                       Second reading


                                                   Mean Value




LIMIT: -


Class
Air particle count
Alert level
Action level
100
> 0.5 µ
76 – 100
> 100
> 5.0 µ
Nil
Nil
100,000
> 0.5 µ
80,000 - 100,000
> 100,000
> 5.0 µ
400 - 560
> 560


REMARKS: -



MICROBIOLOGIST

CHECKED BY
DATE
DATE