    SOP for Carrying Out Air Borne Particle Count in Static Condition

    Standard operating procedure to count non viable particles in classified area using particle counter.

    1.0 OBJECTIVE

    To lay down the procedure for carrying out airborne particle count in the static condition.

    2.0 SCOPE

    This SOP shall be applicable for clean rooms, corridors, laminar air flow and process equipment.

    3.0 RESPONSIBILITY

    Officer /Executive/ Assistant Manager

    4.0 ACCOUNTABILITY

    HOD engineering

    5.0 PROCEDURE

    5.1 Ensure that the area is cleaned and air handling unit is kept ON for at least two hours before taking particle count.
    5.2 The particle count to be performed by an outside agency.
    5.3 Ensure that the particle counter used is duly calibrated and the certificate for the same must be traceable to national/ international standards.
    5.4 Sample 1 cubic feet of air per minute under normalized mode.
    5.5 Calculation for the number of locations for particle count
    5.5.1 Number of sample locations shall be taken as the square root of the area in meter square of clean room/zone.
    5.5.2 This number shall be rounded up to a nearest higher number.
    5.5.3 If the number of location is one then minimum three samples shall be taken at that location.
    5.6 The locations for taking the samples shall be near doors, return air raisers, at corners and near equipment.
    5.7 The count shall be always taken for 1 minute and at working height of about 1 meters from the ground level.
    5.8 After completion of the particle count in a particular room/ zone take the printout.
    5.9.1 The particle count for particle size 0.5 are printed in cubic feet. The printout shows the sampled quantity of air as 1.000 cfm.
    5.9.2 Convert particle count obtained in cubic feet to the particle in cubic meter by multiplying with factor 35.31.
    5.10 Calculate for mean, standard deviation and 95% upper confidence limit, as follows
    5.10.1 Computation of average particle concentration at a location (P)
              ( T1 + T2 + ……… + TX )
    P = -----------------------------------
                                   X
    Where,
    P: Average particle concentration at one location
    T1 to TX: Particle concentration of individual samples
    X: Number of samples taken at the location
    5.10.2 Computation of mean of the averages ( M )
              ( P1 + P2 + ……… + PY )
    M = ----------------------------------
                                   Y
    Where,
    M: Mean of averages
    P1 to PY: individual location averages
    Y: Number of individual location averages
    5.10.3 Computation of standard deviation of the mean ( SD )
                 (P1 – M) 2 + (P2 – M) 2 + …….. + (PY – M )2
    SD = √ ----------------------------------------------------
                                          ( Y – 1 )
    Where,
    SD : Standard deviation of the mean
    5.10.4 Computation of 95% upper confidence limit ( UCL )
                                        SD
    95% UCL = M + ( t X ------)
                                        √Y
    Where t factor for 95% upper confidence limit is as given in table
    Number of individual averages ( Y )
    2
    3
    4
    5
    6
    7- 9
    t
    6.3
    2.9
    2.4
    2.1
    2.0
    1.9
    5.11 If the 95% UCL value exceeds the class limit, take the count at additional evenly distributed locations and compute the 95% UCL for all locations which should not exceed the desired class limit.
    5.12 When only a single location is sampled or when more than nine are sampled computing the mean, standard deviation and 95 % upper confidence limit is not applicable.

    5.13 The frequency for taking the particle count
    5.13.1 For clean rooms, corridors and process equipment shall be once in six months.
    5.13.2 For laminar air flow shall be once in six months.
    5.13.3 The particle count testing shall be carried out as per schedule.
    5.14 Acceptance Criteria
    5.15.1 The cumulative particle concentration for the area should not exceed the desired class limit as given below
    Specified By
    Class
    0.5μ
    5.0μ
    Cubic meter
    Cubic feet
    Cubic meter
    Cubic feet

    ISO 14644
    Class - 5
    3,520
    100
    29
    0
    Class - 7
    3,52,000
    10,000
    2,930
    70
    Class - 8
    35,20,000
    1,00,000
    29,300
    700

    In - house
    100
    3,500
    100
    0
    0
    10,000
    3,50,000
    10,000
    2,000
    57.14
    1,00,000
    35,00,000
    1,00,000
    20,000
    571.43
    5.15.2 The 95% UCL of particle concentration for the area should not exceed the desired class limit.
    5.15 Process equipment having air handling unit shall meet the Class 10,000 requirements.
    5.16 The particle count should meet the In-house specified limits.

    6.0 ABBREVIATION

    6.1 SOP: Standard Operating Procedure
    Also see: HVAC System Validation Tests
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App