    SOP for Cleaning of Equipments in Microbiology Section

    Standard operating procedure to clean the incubators, autoclaves, refrigerator in Microbiology section.

    1.0 OBJECTIVE

    To lay down a procedure for cleaning in microbiology section.

    2.0 SCOPE

    This SOP shall be applicable for maintaining cleaning of Microbiology Section

    3.0 RESPONSIBILITY

    Microbiologist

    4.0 ACCOUNTABILITY

    Head QA & QC

    5.0 PROCEDURE

    5.1 Following group of disinfectants shall be used for cleaning.
    5.2 Clean the ceilings, wall corner, furniture, exhausts grills and blower grills by using dry lint-free cloth or vacuum cleaner.

    5.3 Incubator cleaning

    Switch off the incubator. Remove all the materials from the incubator. Clean the inner area with a disinfectant solution using a lint-free cloth. Make the area completely dry with lint-free cloth. After complete drying, switch on the incubators & can be used for further use.

    5.4 Cleaning of autoclave

    5.4.1 For vertical autoclave switch off the autoclave and remove all the materials from it. Remove the used water wipe off all the inside area with the disinfectant solution.
    5.4.2 Clean the sensor rod with a lint-free cloth. After cleaning, add distilled water containing 10 ml disinfectant solution sufficient to carry out empty run (coil should remain dipped in water/ disinfectant solution even after autoclaving is over.)
    5.4.3 Switch on the autoclave. Carry out the autoclave cycle at 15 lbs at 121 C for 15 minutes, after which switched off autoclave.
    5.4.4 Allow pressure to come down zero, allow it to cool. After cooling remove the disinfectant solution and wipe off the inside area. Use such an autoclave for further use.
    5.4.5 For horizontal autoclave switch off the autoclave and remove all the materials from it. Wipe off all the inside and outside area with the disinfectant solution.
    5.4.6 Clean the sensor rod with a lint-free cloth.

    5.5 Refrigerator

    5.5.1 Remove all the cultures and other materials from the refrigerator and shift it to another refrigerator.
    5.5.2 Allow it to either defrost or switch off the refrigerator. Wipe off the 70% IPA from all the compartments with a clean lint-free cloth.
    5.5.3 Make it dry & then switch on for further use. Transfer all the cultures & materials in it at their original position.

    5.6 Frequency

    Floor cleaning – Twice in a day
    Instruments- Weekly
    Dustbins cleaning –Daily
    Furniture, Walls, Doors, Ceiling-Weekly
    Tube light covers. – Monthly

    6.0 ABBREVIATIONS

    SOP: Standard Operating Procedure
    QA: Quality Assurance
    QC: Quality Control
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App