    SOP for Milliflex Water Filtration Unit

    Standard operating procedure of water filtration unit used to treat the water for analysis.

    1.0 PURPOSE

    To lay down the procedure for operation of Milliflex water filtration unit.

    2.0 SCOPE

    It is applicable to Microbiology lab.

    3.0 RESPONSIBILITY

    Microbiology personnel

    4.0 ACCOUNTABILITY

    Head of Department

    5.0 PROCEDURE

    5.1 Pre-operational checks
    5.1.1 Check the calibration status of the instrument
    5.1.2 Check the status of the sanitation cycle.
    5.1.3 Check the status of the monthly calibration.
    5.1.4 Ensure the head, funnel and stainless steel grid are autoclaved.
    5.1.5 Check the cleaning of the instrument.
    5.1.6 Use sterile membrane filter only.
    5.1.7 Ensure the printer is attached properly.
    5.2 Operation of milliflex
    5.2.1  Install the pump head. Switch on the pump by pressing the On/Off button on the keypad the display shows
                           MILLIFLEX PLUS
                           SERIAL NO. 1234
    5.2.2  Set the date with arrow key by Up and Down
                                    DATE
    5.2.3  After setting the date, set time with arrow key by Up and Down
                                    TIME
    5.2.4  Manual will appear on the screen
                                 MANUAL
    5.2.5  Press OK, tare will appear.
                                   TARE
    5.2.6  Select the sampling option with the help of arrow key
    5.2.7  Enabling data export.
    5.2.7.1  In the main menu, use the down arrow to scroll down to export data, press OK, the display shows
                                  START
    5.2.7.2  Press OK, the display will show
                              OPERATOR
    5.2.7.3  Press up or down keys the display will show
    ABCDEFGHIJKLMNOPQRSTUVWXYZ0123456789
    5.2.7.4  Enter the name by pressing the up or down arrow to highlight a character and then press OK to select it. The selected character is displayed.
    5.2.7.5  Press the up and down arrow to highlight a character selection screen again
    5.2.7.6  Repeat this process for each character. The characters selected so far are displayed when OK is pressed. To delete the last character entered, press C
    5.2.7.7  After entering operator press OK, the display will show
                              CASSETTE
    5.2.7.8  Press OK, the display will show
                                LOT NO.
    5.2.7.9  Enter the media lot No. as described in point No. 5.7.1 to 5.7.6
                             Increment start 1
    5.2.7.10  Enter the date and time, using the up and down arrows, press OK to confirm the completed entry, in each case. The display show.
                                 Begin
    Press start
    5.2.10.11  Press START. The settings just entered will be printed

    5.3  Set up for manual sampling

    5.3.1  In the main menu, use the up or down arrows to select sampling and press OK. The display shows.
                                MANUAL
    5.3.2  Press OK. The display shows:
                            VOLUME OFF
    5.3.3  Select the ON option by arrow key
                            VOLUME ON
    5.3.4  Press OK. The display shows
                           MILLIFLEX 100
    5.3.5  Select the milliflex 250 funnel, press the down arrow, the display shows
                           MILLIFLEX 250
    5.3.6  Press OK, the display shows:
                           DENSITY (g/ml)
                                 1.00
    5.3.7  Use the up or down arrows to select the density of the liquid to be tested. Press OK the display shows.
                             MANUAL
                          MIL 250/1.00

    5.4  Set up for auto sampling

    5.4.1  In the main menu, use the up or down arrows to select sampling and press OK. The display shows.
                             MANUAL
    5.4.2  Press the down arrow the once. The display shows
                      AUTO SAMPLING
    5.4.3  Press OK. The display shows the name of the last auto sampling program selected
    5.4.4  Use the up or down arrows to move through the list of programs to the required program. Press OK. The display shows
                            SELECT
    5.4.5  To proceed with the selected program, press OK. The display shows the main program screen for the selected program. For example
                          MF X 100
                   MIL 100 / 100 ML
    5.5  Use sterile forceps to place the sterile membrane filter on the pump head. Firmly press the sterile funnel onto the pump head.
    Note:  Make sure that the base of the funnel is touch with the top of the pump head filter support.
    5.6  Pour sample into the funnel, slowly to achieve the required volume
    5.7  In Manual mode, if Volume ‘ON’ has been selected, the display will show the volume filled.
    5.8  In Auto Sampling mode, the buzzer will sound when the required volume has been reached.
    5.9  When the required volume has been reached, stop pouring.
    5.10  In Manual mode, press START to start the pump.
    5.11  In Auto Sampling mode, filtration starts automatically.
    5.12  In Manual mode, operate the pump until the sample has been filtered, then press START to dry out the pump head. Wait for the pump to stop.
    5.13  In Auto Sampling mode, when the sample has been filtered, the pump switches automatically to drying. Wait for the pump to stop.
    5.14  Remove the funnel from the pump head and remove the membrane with the sterile forceps.
    5.15  Place the sterile membrane filter on the pre-incubated R2A agar plates. Ensure that there is no air trap inside the filter paper
    5.16  Incubates the plates as required.
    5.17  Before taking next sample wipe with the head with a sporicidal agent and change the SS grid.

    5.18  Calibration

    5.18.1  Enter the configuration menu as described in the configuration section. Press the down arrow until the display shows validation.
    5.18.2  Press OK. The display shows
                           WEIGHT
                      CALIBRATION
    5.18.3  Press OK. The display shows
                  ENTER CODE OR
                           PRESS C
    5.18.4  Enter the security code (press start three times) and then press OK
    5.18.5  When the security code has been entered, the display shows
                   CALIBRATION DATE
                          DD MM YY
    5.18.6  Press the UP or down arrows to enter the calibration date, and then press OK the display shows
                     REMOVE HEAD
                      PRESS START
    5.18.7  Remove the pump head and press START, the display shows
                     MEASURING . . .
                     PLEASE WAIT . . .
    5.18.8  After a few seconds the display shows
                        PUT HEAD
                     PRESS START
    5.18.9  Reinstall the pump head and press START the display shows,
                        PUT 50 G
                 PRESS START 0, 0
    5.18.10  Place the 50 g calibration weight on the pump head and press start
    5.18.11  The pump calibrates with the 50 g weight. Next, the display prompts the user for the 100 g weight and then for the 200 g weight. Follow the prompts as they are displayed. After calibration with the three weights, the display shows,
              SAVE CALIBRATION: OK
                      CANCEL: C
    5.18.12  To proceed and confirm the calibration, press OK. The displays,
                 VERIFICATION
                     PRESS OK
    5.18.13  Follow the prompts displayed to verify the new calibration. After each weight measurement, the display shows the measured value. Press START to confirm the displayed value and proceed to the next step.
    5.18.14  When the verification procedure is completed, The display shows
                  PRINT PRESS START
                      CONTINUE: OK
    5.18.15  Press START, the printer prints the calibration data while the display shows
                        PRINTING
    5.18.16  When printing has finished, the display returns to
                         WEIGHT
                   CALIBRATION

    Frequency of calibration: Monthly


    5.19  Sanitizing

    5.19.1  Sanitize the pump interior, using the automated procedure described in the Sanitization section.
    Sanitizing agents: Disinfectant available for the day of use.
    5.19.2  At the main program screen press OK to enter the main menu the display shows.
                             TARE
    5.19.3  Press the down arrow until the display shows
                        SANITIZING
    5.19.4  Press OK the display shows.
                   PUT EXPANDABLE
                       PRESS START
    5.19.5  Place the sterile membrane filter and a milliflex funnel, on the pump head and press START. The display shows
               FILL 100 ML OF AGENT
                       PRESS START
    5.19.6  Fill the funnel with 100 ml of the sanitizing agent and press START. The display shows
                       SANITIZING
    5.19.7  The pump pulls through 50 ml of the agent, pause for ten minutes, and then pulls through the remaining 50 ml A bar graph at the bottom of the display shows the progress of the operation.
    5.19.8  After the pump has emptied the funnel, its runs a drying cycle and then display shows
               FILL 4 X 100 ML OF WATER, PRESS START
    5.19.9  Fill the funnel with 100 ml of sterile water and press START. The display shows
                          RINSING
    5.19.10  The pump empties the funnel and runs a drying cycle then the display shows
                FILL 100 ML OF WATER
                       PRESS START
    5.19.11  Fill the funnel with 100 ml of sterile water and press START. The rinsing cycle is performed as before. Then the display prompts for the funnel to be filled again, for each rinse, for a total of four rinses when the fourth rinse has been completed, the display shows
                 SANITIZING FINISHED
                        PRESS START
    5.19.12  Press START, remove the funnel and discard it, the display shows
                         SANITIZING
    5.19.13  Press C to go back to the main program screen

    Frequency of sanitization: Weekly

    5.20  Cleaning

    5.20.1  Clean the exterior of the pump with a lint-free mop with 70 % filtered v/v isopropyl alcohol.
    Frequency: Before and after every use.

    5.21  Tare procedure

    Note: The tare procedure adjusts the measurement system in the pump for the weight of the installed pump head. The tare procedure shall be performed whenever the pump head is changed, and whenever the display does not show ’O’, with an empty funnel on the head.
    5.21.1  When the display shows tare, press OK. The displays show
                 REMOVE EXPANDABLE
                         PRESS START
    5.21.2  Remove anything on the pump head and press START. The display shows
                       MEASURING . . .
                       PLEASE WAIT . . .
    5.21.3  After a few seconds, the display returns to the main program screen for the previously selected program.
    5.22  Autoclave the pump head, SS grit before first use.

    6.0  ABBREVIATIONS

    6.1  SOP - Standard Operating Procedure
    6.2  G - Gram
    6.3  Ml - Milliliter
    6.4  % - Percentage
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses


    Follow Pharmaguideline



    DOCUMENTS

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App