Calibration of Disintegration Test Apparatus | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    Calibration of Disintegration Test Apparatus

    Learn how to calibrate the Disintegration Test Apparatus in Pharmaceuticals.

    A) Number of Cycles (With a constant frequency of 29 to 32) per minute:

    1. Record the frequency of moving up and down of the Basket rack assembly, in a given time as shown below.
    Start Time (Mins.)
    End Time (Mins.)
    Side A (Left hand side of the operator).
    Side B (Right hand side of the operator).
    No. Of cycles
    (A)
    No. of cycles per min.
    (A / 5)
    No. Of cycles
    (B)
    No. Cycles per min.
    (B / 5)
    0
    5
    15
    20
    30
    35
    45
    50
    60
    65
    2. Record the frequency (of moving up and down) manually with respect to time.
    3. Do not stop the instrument in between the operation.
    4. Measure the frequency initially and after each fifteen minutes interval from the start, for 5 minutes each.

    5. Record five readings.
    6. Divide the observed reading by five & note as the frequency of moving up & down per minute.
    7. Perform the test for both the basket rack assembly positions (side A & side B) individually.
    Acceptance criteria: The No. of cycles per minute should be within 29 to 32 throughout the period of operation.

    B) Calibration for Temperature Maintenance

    1. Set the temperature to 37°C.
    2. Fill the beaker with water.
    3. Attach the basket rack assembly & start the constant frequency of moving up & down.
    4. Insert the calibrated thermometer in one of the tubes of the basket rack assembly.
    5. Wait till temperature reaches between 36°C to 38°C.
    6. Start recording the readings.
    7. Record the temperature readings initially & after each 15 minutes interval up to 120 minutes from the start.
    8. Perform the test for both the basket rack assembly positions (side A & side B) individually.
    9. While performing the test, do not keep thermometer inside the basket rack assembly constantly, but insert 2 to 3 minutes prior to the measurement to give stable reading.
    Time (In Mins.)
    Temperature (In °C)
    Side A (Left hand side of the operator).
    Side B (Right hand side of the operator).
    Initial
    After 15
    After 30
    After 45
    After 60
    After 75
    After 90
    After 105
    After 120
    Acceptance criteria: The temperature should remain within 37 ± 1°C, throughout the operation.

    C) Calibration of the Distance Traveled

    1. Attach the basket rack assembly firmly.
    2. Attach white paper firmly without kinks on the instrument, parallel to the path of the arm of basket rack assembly.
    3. Select one point on the horizontal arm of the assembly & mark the same on the paper (pointed marker or pen can be used) when the assembly is not moving & at its highest position.
    4. Start the instrument by pressing START / STOP key & followed by pressing the respective timer key.
    5. As soon as assembly reaches the lowest position, mark the same point again on the paper (while doing this activity take the time to decide the exact lowest position & then mark).
    6. Measure the distance between two marks using a calibrated ruler.
    7. Remove the paper.
    8. Perform the test initially, after 60 minutes from the start & after 120 minutes from the start.
    9. Perform the test for both the basket rack assembly positions (side A & side B) individually.
    Acceptance criteria - Distance traveled is between 53 to 57mm.
    Time in mins.

    Distance traveled by the basket

    Side A (Left hand side of the operator)
    Side B (Right hand side of the operator)
    Initial
    After 60
    After 120
    Frequency: Once in a Month.

    Maintenance/ Repair

    When the instrument does not comply with the requirements specified above; the instrument should be labeled as “Out of Calibration” and should be repaired/ serviced. After repairing/ servicing, calibrate the instrument before use.

    REFERENCE

    United State Pharmacopeia
    British Pharmacopoeia

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App