    SOP for Refractometer

    Standard operating procedure of Refractometer to determine the refractive index of liquid samples.

    1.0 OBJECTIVE

    To describe the procedure for operation and calibration of Refractometer.

    2.0 SCOPE

    This SOP is applicable for operation and calibration of Refractometer.

    3.0 RESPONSIBILITY

    Officer/ Executive - Quality Control

    4.0 ACCOUNTABILITY

    Manager - Quality Control

    5.0 PROCEDURE

    5.1 Operations

    5.1.1 Ensure that the instrument is free from dust.
    5.1.2 Ensure that all the parts including eye lens, prisms and mirror are clean before the operation.
    5.1.3 Place the instrument on a table where adequate light can be reflected through the mirror.
    5.1.4 Open the prism lock and place few drops of the liquid testing sample at the temperature specified.
    5.1.5 Allow the liquid to smear on the lower prism surface quickly close the prism lock.
    5.1.6 Adjust the eyepiece cross wire with a knob by rotating it in a clockwise or anti-clockwise direction so that the sharp demarcation line passes through its center.
    5.1.7 Sharp demarcation is produced in the form of a circle having half light & half dark portion.
    5.1.8 A knob provided just above the prism can do fine adjustment.
    5.1.9 Note down the reading from the scale with the help of the magnifier lens of the instrument.
    5.1.10 Use thermometer and water jacket with tubes provided in the instrument whenever it is required.
    5.1.11 Clean the instrument and keep under box when not in use.

    5.2 Calibration

    5.2.1 To achieve accuracy, the instrument should be calibrated against distilled water, which has a refractive index of 1.3325 at 25 °C against the reference liquids given below:
    Carbon Tetrachloride: 1.4603
    Toluene: 1.4969
    Water: 1.3325
    5.2.2 Frequency of calibration: Monthly.

    5.3 Precautions

    5.3.1 Proper handling of the instrument.
    5.3.2 Temperature should be carefully adjusted and maintained since refractive index varies significantly with temperature.

    6.0 ABBREVIATION

    6.1 SOP - Standard Operating Procedure
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App