SOP for UV Cabinet | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    SOP for UV Cabinet

    Standard operating procedure of UV Cabinet used for TLC analysis.

    1.0 OBJECTIVE

    To describe a procedure for operation of UV cabinet.

    2.0 SCOPE

    This SOP is applicable for operation of UV cabinet.

    3.0 RESPONSIBILITY

    Officer/ Executive - Quality Control

    4.0 ACCOUNTABILITY

    Manager - Quality Control

    5.0 PROCEDURE

    5.1 Operation

    5.1.1 Ensure the area is clean.
    5.1.2 Switch on the main power supply.
    5.1.3 Switch on “VISIBLE LIGHT” marked switch.
    5.1.4 Open the door.
    5.1.5 Keep the TLC plate in the chamber.
    5.1.6 Switch off “VISIBLE LIGHT”.
    5.1.7 Close the door.
    5.1.8 Switch on the required UV Light (i.e. Shortwave (254 nm), long wave (360 nm).
    5.1.9 Visualize the sample from outside the chamber through glass provided to visualize the sample inside the chamber.
    5.1.10 After visualizing the sample note down the observation, switch off the UV light and remove the plate from the chamber.
    5.1.11 Take care that any part of the body should not come in direct contact with UV light.

    5.2 Calibration

    5.2.1 Prepare the following standard solutions.
    a) 0.2% w/v Sodium Salicylate in Ethyl Alcohol
    b) 0.04% w/v Sodium Salicylate in Ethyl Alcohol
    5.2.2 Take about 10x10 cm TLC plate of Silica gel G. Apply 5 microlitre, solution a) and solution b) in such a way that the diameter of the spots is about 5 mm and the distance between the spots is about 3cm.
    5.2.3 Allow the spots to dry.
    5.2.4 Observe the spots under 254 nm and 366 nm wavelength radiations and record.
    (NOTE: - the distance between TLC calibration plate and lamp should be identical to test TLC while observation).

    5.2.5 ACCEPTANCE CRITERIA
    5.2.5.1 Spot of Conc. 0.04% w/v should be clearly visible under 254 nm radiations
    5.2.5.2 Spot of Conc. 0.2% should be clearly visible under 366 nm radiations.
    5.2.6 Frequency: Quarterly

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure
    6.2 UV - Ultra Violet
    6.3 TLC - Thin Layer Chromatography
    6.4 nm - Nanometer
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App