Calibration of Viscometer | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    Calibration of Viscometer

    Learn how to calibrate the Viscometer using different viscosity standards.

    Calibration Procedure

    1. Ensure that instrument is ready for use.
    2. Calibrate the instrument by using different calibration standards. (i.e. 50, 500, 1000 and 5000 cps, Use “BROOKFIELD CALIBRATION FLUID”)
    3. Select the spindle as per requirement, by pressing the “SELECT SPINDLE” key, on the display “S” line will begin to blink for four seconds.
    4. Press “UP or DOWN” (­↑↓) arrow keys while blinking, two digit character values will change accordingly. (UP for increase in spindle no. and DOWN for Decrease in spindle no.) Release the arrow key to halt further scrolling and re-press the “SELECT SPINDLE” key.

    5. Select the speed as per requirement, by pressing the “SET SPEED” key, on the display “0.0” line will begin to blink for four seconds.
    6. Select the rpm speed by pressing the “UP or DOWN” (­↑↓) arrow key, the current rpm speed will display on the bottom line of the screen and RPM will blink. [UP for increase the RPM speed & DOWN for Decrease the RPM speed.] Re-press the “SET SPEED” key.
    7. Attach the selected spindle to the lower shaft, lift the shaft slightly, holding it firmly with one hand while screwing the spindle on with the other hand.
    8. Put the test solution beaker below the spindle center and insert the spindle until the fluid level is at immersion groove in the spindle shaft middle of the “SPINDLE MARK LINE”. (Ensure that no air bubble is trapped to the surface of the spindle).
    9. Press the “MOTOR ON/ OFF” key to run the test.
    10. Allow time to stabiles the reading, (The time required for the stabilization is depend on the speed of running & characteristics of the sample fluid.)
    11. Record the stabilized reading in CPS unit in calibration log.
    12. Press the “MOTOR ON/ OFF” key to stop the test and put ‘off’ the instrument mains.
    13. Clean the spindle & instrument after use.
    14. Affix the calibration label on the instrument.
    Related: Determination of Viscosity by Using Ostwald-type, Suspended-level and Rotating Viscometer

    CALIBRATION LOG
    Instrument:____________                                                               Code No:______________
    Make       :_____________                                                               Model:   _______________

    Sr. No.
    Viscosity Standard
    Lot. No.
    Set Spindle No.
    Set
    RPM
    Temp.
    Observed  Viscosity Value in cps
    (Label claim+ 1.0 % CPS)
    1
    50 cps
    2
    100 cps
    3
    500 cps
    4
    1000 cps
    5
    5000 cps
    Calibration Status         :  Satisfactory / Not Satisfactory
    Next Calibration Due    :
    Calibrated By   :                                                                                   Checked By     :
    Date                 :                                                                                   Date                :  

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App