    SOP for Horizontal Autoclave

    Standard operating procedure of horizontal Autoclave used for sterilization of articles used in microbial analysis.

    1.0 OBJECTIVE

    To ensure that the autoclave functions satisfactorily and completes the sterilization cycle effectively.

    2.0 SCOPE

    This SOP is applicable for the horizontal autoclave in the Quality Control Department.

    3.0 RESPONSIBILITY

    Microbiologist - Quality Control

    4.0 ACCOUNTABILITY

    Manager - Quality Control

    5.0 PROCEDURE

    5.1 Procedure for General Cleaning

    5.1.1 Ensure that power supply to the autoclave is switched ‘OFF’ before cleaning. Remove the plug from the socket.
    5.1.2 Clean the outer surfaces of the autoclave with a clean cloth every day.
    5.1.3 To clean the chamber, remove all the remaining materials from the autoclave.
    5.1.4 After the chamber is emptied, pour additional water in the chamber (all around) so that the residual matters/ particles are flushed out, taking care that the particles are not big enough to block the valve.

    5.2 Procedure for Operation

    5.2.1 Switch ON the MCB
    5.2.2 Open the supply of water and compressed air.
    5.2.3 Switch on the Main switch and control switch of the autoclave.
    5.2.4 Open the door of the sterilizer.
    5.2.5 Load the Autoclave.
    5.2.6 Close the door of sterilizer by pressing NST DOOR CLOSE push button.
    5.2.7 DOOR PRECONDITION indicating light will flash ON.
    5.2.8 Switch ON the heater.
    5.2.9 Switch ON the auto-printer
    5.2.10 Switch ON the temperature and pressure scanner.
    5.2.11 Press F4 button on the control panel
    5.2.12 Press F1 to check or set the parameter of desired process OR press F4 to go to the main menu.
    5.2.13 Press F2 will show the START SCREEN
    5.2.14 Select the desired process by pressing F3 soft key.
    5.2.15 Press F2 for the parameter of the desired cycle to print
    5.2.16 Select Auto mode by pointing AUTO/MANUAL knob towards AUTO
    5.2.17 Press F1 to start the desired process. Operation cycle will start.
    5.2.18 After completion of cycle switch off the heater.
    5.2.19 Let the pressure of the chamber to become Zero.
    5.2.20 Again select MANUAL by pointing AUTO/MANUAL knob towards Manual.
    5.2.21 Push NST DOOR OPEN Push button to open the door.
    5.2.22 Unload the autoclave.
    5.2.23 Switch off all the switch and MCB.
    5.2.24 Maintain the usage record.

    5.3 Precautions

    5.3.1 Use thermal gloves to open the autoclave.
    5.3.2 Do not touch the autoclave when it is hot.
    Related: Autoclave Validation in Pharmaceuticals

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure
    6.2 SCDM - Soybean Casein Digest Medium
    6.3 LAF - Laminar Air Flow
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App