    SOP for Calibration of Thermometer

    Standard operating procedure of calibration of thermometer using standard thermometer certified by Approved Laboratory.

    1.0 OBJECTIVE

    To describe the procedure for calibration of the thermometer.

    2.0 SCOPE

    This SOP is applicable for calibration of the thermometer.

    3.0 RESPONSIBILITY

    Quality control executive

    4.0 ACCOUNTABILITY

    Manager – Quality control

    5.0 PROCEDURE

    5.1 Calibration Procedure

    5.1.1 The thermometer for calibration shall be mercury in glass type.
    5.1.2 Take a 1000 ml beaker filled with about 850 ml of Silicon oil. Use water for (0–0°C) and silicon Oil for the temperature above 50°C.
    5.1.3 Put the beaker in an oil bath fitted with a stirrer. The stirrer shall be dipped in the beaker leaving about 2 cm space from the bottom of the beaker.
    5.1.4 Immerse the standard thermometer certified by Approved Laboratory, into the silicon oil vertically approx. 8.0 cm leaving the remainder of the stem and the upper expansion chamber exposed to ambient temperature. Clamp tightly.
    5.1.5 Immerse the thermometer to be calibrated into the silicon oil vertically. The mercury bulb of the thermometer shall be placed at the same level of the mercury bulb of the standard thermometer. Clamp tightly.
    5.1.6 In case of minimum/maximum thermometers, at the time of calibration, ensure that both the arm should show the same temperature.
    5.1.7 Start the stirrer taking precaution that it does not touch the thermometers.
    5.1.8 Start heating the oil bath at the rate of 2°C per minute and compare the temperature observed at different increments (Annexure - I).
    5.1.9 The temperature difference shall not be more than ±2°C. Any thermometer found beyond the acceptable limit shall be discarded. .

    5.2 Identification Number

    5.2.1 Identification number shall have 8 characters.
    5.2.2 The first two character shall always be T and C respectively.
    5.2.3 The third character shall be dash (-).
    5.2.4 The fourth & fifth character shall be numeric and denote original number.
    5.2.5 The sixth character shall be slash (/).
    5.2.6 The seventh & eighth character shall be numeric and shall denote number of re-calibration.
    5.2.7 A typical example shall be as follows: TC-01/01
    5.2.8 The thermometer shall have an identification number and standardization date labeled on it.
    5.2.9 All thermometers shall be standardized once a year and calibration record maintained (Annexure - II).
    5.2.10 Standard thermometers shall be sent for re-standardization based on the calibration certificate given by the Approved Laboratory.
    5.2.11 Alternatively calibrated thermometer can also be used.

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure
    6.2 No. - Number
    6.3 QA - Quality Assurance
    6.4 NA - Not Applicable
    6.5 CCF - Change control format
    6.6 QC - Quality Control
    6.7 R&D - Research and Development

    ANNEXURE – I
    S. NO.
     RANGE OF THERMOMETER
    SET TEMP. FOR STANDARDISATION

    1

    - 10°C to 50°C
    i)  12°C 
    ii)  25°C
    iii) 38°C
    iv) 50°C    

    2

    - 10°C to 110°C
    i)   25°C
    ii)  50°C  
    iii) 80°C 
    iv) 110°C   

    3

    - 10°C to 250°C
    i)    60°C
    ii)  120°C
    iii) 180°C
    iv) 220°C

    4

    - 10°C to 360°C
    i)    60°C
    ii)  120°C
    iii) 180°C
    iv) 250°C

    Note:  If a pre-certified/calibrated thermometer is used the range given in the certificate shall be used for standardization instead of above set temperatures.

    ANNEXURE - II
    STANDARDISATION RECORD
    NAME OF EQUIPMENT                     :  THERMOMETER
    THERMAMETER Sr.No                      :
    FREQUENCY                                      :  YEARLY

    IDENTIFICATION NO. OF STD. : 
    THERMOMETER

    DATE OF STANDARDISATION
    THERMOMETER NO.
    SET
    TEMP.
    STANDARD
    THERMOMETER
    OBSERVED GAUGE IN MM
    THERMOMETER TO BE STANDARDISED
    REMARKS
    CALIBRATED BY
    CHECKED
    BY
    CORRECTION
    TO READING
    ACTUAL
    TEMP.
    TEMP.
    OBSERVED
    CORRECTION
    TO READING






    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App