    SOP for Calibration of Coulter Counter

    Standard operating procedure to calibrate the coulter counter.

    1.0 OBJECTIVE

    To calibrate the equipment for reliable and accurate results.

    2.0 SCOPE

    This procedure is applicable to calibrate the coulter counter installed in the quality control department.

    3.0 RESPONSIBILITY

    3.1 Doing: Technical Assistant
    3.2 Checking: Executive /Manager

    4.0 ACCOUNTABILITY

    Head of the Department

    5.0 PROCEDURE

    Frequency: Once in a three month
    5.1 Program and operate the instrument as per SOP.
    5.2 Use the ‘menu cursor’ keys to select the following option on the ‘SET UP’ menu.
    5.3 Set the parameters as follows.
    Orifice size :70 & 100 µ alternately
    Diameter/Length: Enter the aperture diameter and length of the orifice tube selected for calibration.
    Set up: Select 'Automatic'
    Analysis: Select 'Sample'
    Calibration: Select 'New'
    Size: Enter the (number mode) of the latex particles used for calibration and the assay Diameter of the chosen latex size should be 20 % of the aperture diameter for optimum results.(e.g. 2.15 E1 means 21.5 m)
    Size units: Select mm
    5.4 Select ‘FULL RANGE’.
    5.4.1 X-Axis ----> Select ‘Lin Dia’.
    5.4.2 Y-Axis ----> Select ‘Number Diff’.
    5.4.3 Blank subtract ---> Select No.
    5.4.4 Channels-----> 256
    5.4.5 Channel data --> No.
    5.5 Observe reading as per S.O.P.
    5.5.1 Move the upper and lower cursor to coincide with peak height.
    5.5.2 Press SET UP key and select record on the calibration line on SET UP menu.
    5.5.3 The new Kd factor will display .
    5.5.4 Press the CAL key, the instrument is now calibrated for that electrolyte solution and orifice tube for diameter and volume.
    5.5.5 Print the graph and set up menu with new Kd factor for recording. Record the observation as per annnexure. If calibration is out of limit, follow the S.O.P.

    6.0 ABBREVIATIONS

    6.1 No.= Number
    6.2 Lin.Dia.= Linear diameter
    6.3 Diff = Difference
    6.4 mm= Micrometer
    6.5 m= Micron

    ANNEXURE-I
    Q. C. DEPARTMENT
    CALIBRATION REPORT OF COULTER COUNTER
    CALIBRATION  DATE
    DATE OF LAST CALIBRATION DONE
    NEXT DUE FOR
    CALIBRATION

    INSTRUMENT DETAILS
    INSTRUMENT NAME
    INSTRUMENT MAKE
    INSTRUMENT
    IDENTIFICATION NO.
    Name of std. Calibration solution  : CC standard L5 Polystyrene Latex
    Lot No.                                        :
    Size of Latex particles                   : 
    Conductivity fluid                          : Filtered N.Saline
    Label claim                                   :
    For 70µm aperture diameter          : 3 drops per 50ml
    For 100µm aperture diameter        : 3 drops per 50ml

    OBSERVATION

    Orifice Diameter     Observed Kd value                                       Limit of  Kd value
    70 microns                                                                                  677+/-10%
    100 microns                                                                                930 +/- 10%
    Orifice Diameter     Observed diameter of latex particles used      Limit of  latex particles
    70 microns                                                                                  +/- 2 % of label claim. (Limit:     to     µm)
    100 microns                                                                                                     
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App