    SOP for Calibration of Karl Fischer Apparatus

    Standard operating procedure to calibrate the Karl Fischer Apparatus with Disodium tartrate.

    1.0 OBJECTIVE

    To calibrate the equipment for reliable and accurate results.

    2.0 SCOPE

    This procedure is applicable to calibrate the Karl Fischer apparatus, installed in the quality control department.

    3.0 RESPONSIBILITY

    3.1 Doing: Technical Assistant
    3.2 Checking: Executive /Manager

    4.0 ACCOUNTABILITY

    Head of the Department

    5.0 PROCEDURE

    Frequency: Once in a month
    5.1 Operate the instrument as per SOP.

    5.2 FOR BURETTE CALIBRATION

    5.2.1 Place the burette outlet in 10ml standard calibrated measuring cylinder
    5.2.2 Press the START key and then press ENTER key. Then again press START key.
    5.2.3 When the display show 3ml, immediately press STOP key.
    5.2.4 Note down the observed volume of Karl Fischer reagent in the measuring cylinder.
    5.2.5 Now repeat step No 5.2.2 to 5.2.4 for 6ml and 9ml.
    5.2.6 Record the obtained volume and displayed volume in the enclosed format.
    5.2.7 The volume displayed on the instrument and the volume collected in cylinder should not vary by more than ± 0.1 ml, otherwise, follow the S.O.P.
    5.2.8 Record the results in Annexure.
    5.2.9 Frequency of calibration: Once in a month.

    Related: Water Content Determination by Karl Fischer

    5.3 FOR INSTRUMENT PERFORMANCE

    5.3.1 Add accurately weighed 100mg of Disodium tartrate into the titration vessel containing neutralized methanol and test for water content.
    5.3.2 Calculate the % water by using the formula given below.
                                           Volume of KFR consumed in ml X 100 X Factor
    % Water (W/W) = --------------------------------------------------------------
                                                      Weight of disodium tartrate in mg
    5.3.3 The performance of the instrument is satisfactory if the difference between the obtained values is within the tolerance limit of 15.66 ± 5% i.e. 14.87% -16.44, otherwise follow the S.O.P.
    5.3.4 Repeat the step 5.3.1 & 5.3.2 and take total five reading.
    5.3.5 Record the results in Annexure-I.
    5.3.6 Frequency of calibration: Once in a month.

    6.0 ABBREVIATIONS

    6.1 % = Percentage
    6.2 mg = milligram
    6.3 W/W = Weight by Weight
    6.4 S.O.P. = Standard operating procedure number
    6.5 ml = milliliter
    6.6 KFR = Karl Fischer reagent

    ANNEXURE-I
    Q. C. DEPARTMENT
    PERFORMANCE CHECK REPORT OF KARL FISCHER APPARATUS
      
    PERFORMANCE CHECK  DATE
    DATE OF LAST PERFORMANCE CHECK DONE
    NEXT DUE FOR
    PERFORMANCE CHECK

    INSTRUMENT DETAIL
    INSTRUMENT NAME
    INSTRUMENT MAKE
    INSTRUMENT
    IDENTIFICATION NO.

    OBSERVATION

    Sr. No.
    Wt. taken of di-sodium   tartrate
    (mg)
    Volume of KFR required
    (ml)
    Factor
    mg. H2O/ml
    % Water obtained
    Tolerance
    1
    14.87 % to 16.44 %
    2
    3
    4
    5
    % RSD :
    (NMT 1.0 %)
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App