SOP for Steam Penitration Study by Bowie-Dick Test | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    SOP for Steam Penitration Study by Bowie-Dick Test

    Standard operating procedure to study the even steam penetration of steam in double door autoclave/sterilizer.

    1.0 Objective 

    To study of steam penetration evenly in double door Autoclave with use of Bowie-dick type test kit.

    2.0 Scope

    This procedure is applicable to study even and rapid steam penetration in Double door Autoclave.

    3.0 Responsibility 

    3.1 Doing : Technical Assistant/Executive
    3.2 Checking : Executive/Manager

    4.0 Accountability 

    Head of the Department

    5.0 Procedure 

    5.1 Keep the Bowie-Dick type test pack in the lower shelf of autoclave near drain on SS stand.
    Set the Bowie-Dick cycle for 18 min.
    5.2 Run the cycle as per SOP for 18 min.
    5.3 After Completion on cycle remove the test pack from chamber, remove the cover and paper sheets Around Test paper.
    5.4 Observe the Bowie-Dick test paper for even colour change of the indictor which is distributed on test paper.
    5.5 Record the results and attach this test paper with annexure.
    5.6 The test should be considered satisfactory if there is a uniform colour change through out the indicator.
    5.7 It there is a failure check the pervious load and previous results.
    5.8 Inform to engineering department for the rectification and corrective action.

    6.0 ABBREVIATION

    NIL

    ANNEXURE-I
                                                                                                                                                                               SOP No : ___________
    MICRO SECTION QUALITY CINTROL

    RESULTS OF BOWIE-DICK TYPE TEST KIT
    (Study of steam penetration study in Double door Autoclave)
    Date of Sterilization
    :
    Ster-Hold Time
    : 8 minutes
    Sterilization Cycle No.
    :
    Autoclave ID No.
    :
    BOWIE-DICK TYPE TEST KIT
    Manufacturer
    :
    Mfg date
    :
    Lot Number
    :
    Exp. date
    :

    Observation :
    Conclusion: Steam penetration is SATISFACTORY / NOT SATISFACTORY
    Acceptance Criteria: The colour of Bowie-dick test paper should be changed from yellow to blue evenly


























    Analyst                                                          Date of report                                        Checked By

    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App