Balances are used to weigh different things as raw material, dispensed material and samples. Sometimes very low quantity is weighed for analysis. An error in the weighed material can cause a big variation in analytical results.
Following circumstances may cause the error in the weighing:
Assay, related substances and other quantitative analytical tests require accurate weighing for better results. Near the lower and upper capacity of the balance, weighing may be inaccurate. Therefore, a weighing range must be defined for all balances.
Related: Calibration of Analytical Balance
If any balance has the operating range from 0.01 mg to 110 gm then it should be used for weighing the material from 0.01X50 = 0.50 mg to 110X80% = 88.00 gm. Detailed guidance on the balances can be learned through OIMLR76-1 Non-automatic Weighing Instruments.
But according to USP <41> Weights and Balances, a simple theoretical calculation can be used to calculate the minimum weight for any balance. To weigh 10 mg sample you have to use 10X0.1% = 0.01 mg least count balance. It shows that if you have a balance with least count 0.01 mg, you can weigh minimum 10 mg sample on it. Other weights can be calculated accordingly.
Update: Simply you can calculate the minimum limit by multiplying least count by 1000.
Minimum Weighing Limit = Least Count x 1000
- Balance for analytical use must be closed; an open balance may cause inaccurate weighing.
- Hygroscopic material gains weight when it comes in contact with the air causing unstable weighing.
- Airflow or current in the balance room may cause a disturbance in accurate weighing.
- Unleveled surface or balance may lead to inaccurate weighing.
- A vibrating surface can also cause inaccurate weighing.

Related: Calibration of Analytical Balance
Weighing Range of Balances:
Some manufacturers direct in their operating manual to weigh minimum 1.0 mg on balance having 0.01 mg least count i.e. least count X 100. But the thumb rule followed in pharmaceutical industries for lower limit is least count X 50 and the upper limit is 80% of the capacity of the balance.If any balance has the operating range from 0.01 mg to 110 gm then it should be used for weighing the material from 0.01X50 = 0.50 mg to 110X80% = 88.00 gm. Detailed guidance on the balances can be learned through OIMLR76-1 Non-automatic Weighing Instruments.
Update: Simply you can calculate the minimum limit by multiplying least count by 1000.
Minimum Weighing Limit = Least Count x 1000
please give us the ISO and Auditing documents of electrical dept also sir...
ReplyDeleteinformative article
ReplyDeleteWhere to find the least count for a specific weighing balance
ReplyDeleteLeast count: The minimum display that can be shown is the least count, this will be issued by the weighing balance vedor as per our URS.
DeleteNo need to perform daily verification because all ranges covered in monthly calibration
ReplyDeleteDaily weighing balance verification should perform due to while performing housekeeping activity balance platform will get disturbed, hence to confirm the balance is in good level for usage balance shall be verfied daily
ReplyDelete