SOP for Performance Check of Autodiluter


Standard operating procedure to calibrate Autodiluter used to dilute the solutions in pharmaceuticals.

1.0  Objective

       To check the performance of Autodiluter.

2.0  SCOPE

       This procedure is applicable to calibrate Autodiluter.

3.0  Responsibility              

3.1  Doing      : Technical Assistant
3.2  Checking : Executive /Manager   

4.0  Accountability

       Head of the Department

5.0  Procedure

5.1  Frequency : once in a 3 months
5.1  Clean the instrument with dry cotton cloth.
5.2  Operate the instrument as per S.O.P.
5.3  For Right syringe (B)
      1. Take five empty beaker, they should be conditioned & weigh previously.
5.3.2  Suck 100µl with right syringe (B) & Note down the water weight. Further take 200µl, 300µl, 400µl & 500µl & Note down the water weight.

5.3.4  Check the linearity as shown in Annexure-I.
5.4  For left syringe (A)
5.4.1  Take five empty beaker, they should be conditioned & weigh previousely.
5.4.2  Suck 1 ml with left syringe (A) & note down the water weight.
5.4.3  Further take 2 ml, 3 ml, 4 ml, & 5 ml and note down the weight.
5.4.4  Check the linearity as shown in Annexure –II.

6.0  ABBREVIATION

6.1  μl = Microliter
6.2  ml = Millilitre


ANNEXURE –I
Q. C. Department  
                                  PERFORMANCE CHECK RECORD FOR AUTODILUTER
Name of Instrument    :
Ref No.     :
Make                          :
Issue Date :
Model No.                  :
Issued By  :
Id. No.                        :


                                                                                                       For Right Syringe (B)
Sr. No.
Volume
Wt. Of water in mg
Linearity
1
100 µl


2
200 µl

3
300 µl

4
400 µl

5
500 µl


         Linearity        :  Limit         :   NLT 99.0%         
         Remarks                           :
         Calibration date                 :    
         Calibration due date           :
         Calibrated by                    :
         Checked  by                     :   

ANNEXURE –II
Q. C. Department

                                                                                                       For Left Syringe (A)
Sr. No.
Volume
Wt. Of water in mg
Linearity
1
1 ml


2
2 ml

3
3 ml

4
4 ml

5
5 ml


         Linearity        :  Limit         :   NLT 99.0%
         Remarks                           :
         Calibration date                 :    
         Calibration due date           :
         Calibrated by                    :
         Checked  by                     :