    SOP for Fumigation of Microbiology Laboratory

    Standard operating procedure of fumigation process in microbiology lab using chemicals and precaution taken during the fumigation.

    1.0 OBJECTIVE

    To lay down the procedure for fumigation of Microbiology Laboratory.

    2.0 SCOPE

    This SOP shall be applicable to Quality Control Dept.

    3.0 RESPONSIBILITY

    Microbiologist

    4.0 ACCOUNTABILITY

    Head Quality Control Department.

    5.0 PROCEDURE

    5.1 Apparatus

    Fumigation machine

    5.2 Reagent

    5 % Gramicid Solution

    5.3 Fumigation Procedure

    Frequency: Once in a week or if the microbial counts exceed the limits in Microbiology Testing area .
    5.3.1 Connect the machine electric plug to the main supply.
    5.3.2 Keep the machine 2 feet above the ground level.
    5.3.3 Open the lid of the tank.
    5.3.4 Add 2 liters of 5 % Gramicid solution and close the lid.
    5.3.5 Adjust the blower at approximately 45° angle.
    5.3.6 Put on the MAIN switch.
    5.3.7 Switch on the blower.
    5.3.8 Fogger starts emitting fine fog.
    5.3.9 Keep the fog generator control minimum, to get fine particles of the gramicid solution.
    5.3.10 Keep the fogger on till it forms sufficient fog in the area and then shift to another area.
    5.3.11 Air handling unit should be put OFF before starting fumigation.
    5.3.12 After fumigation keep all the doors closed.
    5.3.13 After completion of fumigation keep the area closed for not less than 6 hrs.
    5.3.14 Switch on the AHU 1 hour before starting the work for de-fumigation
    5.3.15 Maintain fumigation Record.
    Related: SOP for Guidelines for Working in Microbiology Laboratory

    5.4 Precautions

    5.4.1 Make sure the area to be fumigated cleaned thoroughly.
    5.4.2 Cover the electronic equipment with a plastic sheet.
    5.4.3 AHU should be put OFF 10 minutes before the fumigation starts.
    5.4.4 AHU can be put ON 1 hour before starting work.
    5.4.5 Use proper safety apparel such as gowns, rubber hand gloves and safety goggles during fumigation of the lab.

    6.0 ABBREVIATIONS

    6.1 SOP: Standard Operating Procedure
    6.2 Dept.: Department
    6.3 AHU: Air Handling Unit
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App