    SOP for Leak Seal Test Apparatus

    Standard operating procedure to operate the leak test apparatus and testing of strips or leakage using the methylene blue.

    1.0 OBJECTIVE

    To lay down the procedure for operating leak seal test apparatus.

    2.0 SCOPE

    This SOP shall be applicable to Quality Control Dept.

    3.0 RESPONSIBILITY

    Supervisor Q.C

    4.0 ACCOUNTABILITY

    Head- Quality Assurance

    5.0 PROCEDURE

    5.1 Ensure the cleanliness of the area and the equipment.
    5.2 Ensure that there shall not be any remnants of previous product handled in the instrument.
    5.3 Fill the desiccator with water just below the level of the sieve, add few drops of methylene blue and mix to give a light blue color.
    5.4 The number of strips/blister taken should be sufficient to cover all the pockets of sealing roller.
    5.5 Take the exact number of strips/blisters and dip them into the water in the desiccator.
    5.6 Place the plastic sieve over the strips and place the lid over the desiccator.
    5.7 Connect the desiccator to the vacuum pump.
    5.8 Switch ‘ON’ the vacuum pump.
    5.9 Apply 15 inches of Hg pressure (negative pressure), Hold the vacuum for 2 minutes after the desired pressure is reached.
    5.10 Switch ‘OFF’ the vacuum pump.
    5.11 Release the pressure slowly (taking at least 30 seconds) and take out the strips.
    5.12 Wipe them to dry with a clean dry lint-free cloth.
    5.13 Defoil the strips/blisters and observe for blue spots/moistness/water inside
    5.14 If blue spots/ moistness or water is found inside the pockets, strips are failing in the test. Inform the results of the test to Q.A Person for necessary action.
    5.15 Take care to discard all tested tablets/Capsules by putting in a container filled with water.
    5.16 Q.C Dept. should perform the test for all the Finished Product batches received from Production dept.

    6.0 ABBREVIATIONS

    6.1 SOP: Standard Operating Procedure
    6.2 QA: Quality Assurance
    6.3 QC: Quality Control
    6.4 Dept.: Department
    6.5 Hg: Mercury
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App