    SOP for Cleaning, Operation and Calibration of Antibiotic Zone Reader

    Standard operating procedure to clean and calibrate the antibiotic zone reader used in antibiotic assay.

    1.0 OBJECTIVE

    To lay down the procedure for cleaning, operation and calibration of Antibiotic Zone Reader.

    2.0 SCOPE

    This SOP shall be applicable to Quality Control Dept.

    3.0 RESPONSIBILITY

    Microbiologist

    4.0 ACCOUNTABILITY

    Head Quality Control Department.

    5.0 PROCEDURE

    5.1 Cleaning

    5.1.1 Clean with a non-shedding duster dipped in the disinfectant solution, the outer surfaces of the antibiotic zone reader daily.
    5.1.2 Record the cleaning of the “Equipment Cleaning Register”.

    5.2 Operation

    5.2.1 Ensure that the instrument is clean and free from dust.
    5.2.2 Connect the main cord to the supply socket and switch it on.
    5.2.3 There is a switch on the right side panel of the instrument. Put this switch on.
    5.2.4 The indicator lamp will be on and the main light source of the zone reader will be on.
    5.2.5 Set the vernier scale to zero-zero coincidence.
    5.2.6 Place the microbiological assay plate and measure accurately the zone of inhibition or exhibition.
    5.2.7 Keep the plate on the stage. Adjust the plate such that one side of the circumference of the zone touches the marked line of the view glass.
    5.2.8 Turn the knob in the clockwise direction so that the plate moves across the marked line of the view glass until the opposite edge of the zone is in tangent to the marked line of the view glass.
    5.2.9 The scale reading gives nearest 0.2 mm on the rotating scale and 1/5 mm on the fixed vernier scale.
    5.2.10 Switch off the mains and keep the instrument under cover when not in use.
    5.2.11 Record the activity in the log book.

    5.3 Calibration

    5.3.1 The instrument is to be calibrated once a month.
    5.3.2 Mark the diameter of the Acrylic disc by drawing the line.
    5.3.3 Measure the diameter of Acrylic discs having the diameter of i)18 mm, ii) 20 mm, iii) 25 mm with the calibrated vernier and note the reading.
    5.3.4 While measuring the diameter, measure along the line of diameter by vernier and by Antibiotic zone reader.
    5.3.5 Switch on the main switch on the zone reader.
    5.3.6 Turn the right side knob in such a way that the fixed vernier scale matches with the rotating scale at ‘0’.
    5.3.7 Keep the earlier measured disc in the Petri plate. Keep the plate on the stage. Adjust the plate such that, point where diameter line intercepts circumference of the disc, touches the marked line of the view glass.
    5.3.8 Rotate the right side knob in the clockwise direction so that the opposite side’s point of interception of the circumference of disc touches marked the line of the view glass.
    5.3.9 Note down the readings from the rotating scale.
    5.3.10 Likewise measure three discs of each of different diameter, which is previously measured with calibrated vernier calipers.
    5.3.11 Find out the difference. The difference should not be more than ± 0.2 mm
    5.3.12 Mention date of calibration, the due date of re-calibration & performed by on label and affix on the instrument.

    5.4 Maintenance

    5.4.1 The instrument should be handled properly.
    5.4.2 The light source should not be kept on for a long period of time.
    5.4.3 The lens from which light is reflected should be clean and free from dust.

    6.0 ABBREVIATIONS

    6.1 QC: Quality Control
    6.2 SOP: Standard Operating Procedure
    6.3 Dept.: Department
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App