SOP for Maintenance and Calibration of Heating Block | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    SOP for Maintenance and Calibration of Heating Block

    Standard operating procedure of maintenance and calibration of heating block used for endotoxin testing.

    1.0 OBJECTIVE

    To lay down a Procedure is to provide guidelines for Operation, Maintenance & Calibration of Heating Block.

    2.0 SCOPE

    This procedure is applicable for Heating Block which is installed in microbiology laboratory of Quality Control Department.

    3.0 RESPONSIBILITY

    Microbiologist

    4.0 ACCOUNTABILITY

    Head QA & QC

    5.0 PROCEDURE

    5.1 Connect the 230V AC power supply to the power connection properly and Switch ON the Mains.
    5.2 Open the lock of control panel situated on the front side of equipment with the help of the key.
    5.3 Switch “ON” the red switch situated at the control panel, which is indicated by glowing of the red lamp.
    5.4 The equipment temperature ranges from 0°C–100°C, so adjust the temperature at 37°C ± 1°C of the equipment by pushing the SET button and rotating clockwise or anti-clockwise temp. set knob of the equipment placed on the control panel.
    5.5 Confirm the temperature that is displayed on the digital display and then releases the SET button.
    5.6 Now the heater will start.
    5.7 After attaining the SET temperature the heater gets automatically “OFF”.
    5.8 Observe again after 30 minutes of attending the set temperature. If the displayed temperature is stable at the set temperature, then the equipment is ready for use.
    5.9 Maintenance/Cleaning: Daily
    5.9.1 Clean the outer surface of the Heating block with a lint-free cloth soaked in a mild soap solution, followed by wiping with a dry lint-free cloth.
    5.10 Maintenance/Cleaning: Monthly
    5.10.1 Switch OFF the Main switch of the equipment and then switch OFF from Main Supply.
    5.10.2 Affix the label “UNDER MAINTENANCE/UNDER CLEANING”.
    5.10.3 Remove the heating blocks and transfer to washing area.
    5.10.4 Wipe out the outside surface of the equipment with liquid detergent and then wipe out the base of the heating block.
    5.10.5 Place the blocks in hot WFI for 30 minutes and then wash with a liquid detergent solution. Use nylon brush if there is any stain.
    5.10.6 Again wash with WFI to remove the detergent traces and allow drying at room temperature.
    5.10.7 After drying fix the blocks in their respective place and remove the label “UNDER MAINTENANCE/UNDER CLEANING”.
    Related: SOP for Heating Block Validation

    5.11 Calibration

    5.11.1 Internal Calibration frequency once in six months & External Calibration frequency once in a year.
    5.11.2 Internal Calibration carried out by Microbiology department and External Calibration carried out by engineering department or approved outside party.

    5.12 Internal Calibration

    5.12.1 Operate the Heating block as per SOP for operating instruction for the heating block.
    5.12.2 Insert five BET tubes in a different location of wells and then add 200 ml SWFI with the help of micropipette.
    5.12.3 Now insert a calibrated thermometer into the tubes containing 200 ml WFI.
    5.12.4 Set the Heating block at 37°C temperature.
    5.12.5 After attaining the heating block temperature, note down the temperature for 1 hour at the interval of 02 minutes for each set temperature points.
    5.12.6 Compile the data and find out the temperature uniformity throughout the 1 hour period. It should be ± 1°C.
    5.12.7 After calibration, immediately put the “CALIBRATION” tag duly filled with necessary information.

    5.13 External Calibration

    5.13.1 Temperature mapping.
    5.13.2 When 37°C temperature achieves, insert thermocouple in the test tube filled with water.
    5.13.3 Connect the thermocouple with the data logger. Record the thermocouple reading for 1 hour at an interval of 1 minute.
    5.13.4 Determine the heat distribution studies by analyzing the temperature profile data.

    5.14 Acceptance criteria

    5.14.1 Temperature recorded by each probe at different locations must be in the range of 37°C ± 1°C.

    6.0 ABBREVIATIONS

    SOP: Standard Operating Procedure
    QA: Quality Assurance
    QC: Quality Control
    NA: Not Applicable
    SS: Stainless steel
    WFI: Water for Injection
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App