SOP for Bacterial Endotoxin (LAL) Test | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    SOP for Bacterial Endotoxin (LAL) Test

    Standard operating procedure of Bacterial Endotoxin detection in raw material and finished products.

    1.0 Purpose

    To lay down the procedure to perform Bacterial Endotoxin (LAL) test for detection of endotoxin in the material to be tested.

    2.0 Scope

    It is applicable to Quality control department.

    3.0 Responsibility

    Microbiologist

    4.0 Accountability

    Head of Department

    5.0 Procedure

    5.1 Precaution to be taken during LAL testing
    5.1.1  Check the calibration status of the triobloc.
    5.1.2  Check the temperature of triobloc before keeping the sample tubes into the wells, it should be 37±1°C
    5.1.3  All ready to use glassware / accessories must have proper certification for endotoxin-free quality.
    5.1.4  Depyrogenated glassware only shall be used for the test.
    5.1.5  Storage and reconstitution of control standard endotoxin and lysate shall be according to manufacturer recommendations.
    5.1.6  All the reagents to be used for the test must be within its expiry date.
    5.1.7  Do not vortex lysate on reconstitution; shake gently to mix it properly.
    5.2  Depyrogenation of glassware
    5.2.1  Depyrogenation of LAL reaction tubes / dilution tubes / accessories
    5.2.1.1  Tear off piece of aluminum foil large enough to be folded and cover the required number of tubes / accessories to be pyrogen free.
    5.2.1.2  Place prepared tubes / accessories on aluminum foil, Wrap foil around tubes / accessories so that they are completely covered.
    5.2.1.3  Put the tubes / accessories in dry heat sterilizers and run as per the SOP.
    5.2.1.4  After depyrogenation is completed unload the tubes / accessories and observe the pack, no tears in foil should be observed. If tears are noted rewrap the tubes / accessories again and put for depyrogenation.
    5.2.1.5  Alternatively depyrogenated tubes / accessories (ready to use) can be directly used.
    5.3  Calculation of Maximum  valid  dilution  (MVD) and Minimum valid concentration  (MVC)
    For Raw material   MVC  =   λ (EU / ml)                                  
                                                   EL (EU / mg)
            
    For Finished products   MVD = EL x Potency of product
                                                             λ          (EU / ml)
    For Finished products   MVD = E.L.C.
                                                            λ
    Whereas,
    EL = Endotoxin limit specified in the individual monograph of product, in EU/ml or EU/mg.
    Potency of Product  = Concentration of product in Units or mg/ml
    λ  Labelled sensitivity of lysate.
    E.L.C.  Endotoxin limit concentration Specified in the monograph, in specified ml of product.
    If Endotoxin limit of the product is not specified or for a new product, calculate Endotoxin limit as follows:
    Endotoxin limit  =    K
                                    M
    K = 5.0 EU/kg for parenteral preparation (0.2 EU/kg. for those drugs administered intrathecally)
    M = Maximum human dose /kg/hour

    5.4  Reconstitution of LAL Reagent and Storage condition.
    5.5.1  Store the lyophilized lysate as mentioned by the manufacturer.
    5.5.2  Reconstitute the lysate with LAL Reagent water (LRW) as per the instruction is given by the manufacturer.
    5.5.3  Reconstituted lysate may be stored as per manufacturer.
    5.5.4  Reconstituted Lysate may be frozen and thawed once only for use (Follow as per the manufacturer instruction)
    5.5  Reconstitution of Controlled Standard Endotoxin (CSE) and storage condition.
    5.5.1  Reconstitute the CSE with LRW as per the manufacturer instruction and vortex it for the given time period.
    5.5.2  Reconstituted CSE may be stored per manufacturer temperature / time period mentioned.
    5.5.3  Vortex the CSE at least for 5 minutes before using and vortex the Endotoxin at least for one
    minute before use and after every ten minutes of standing.
    5.6  Confirmation of Labelled Sensitivity of LAL Reagent.
    5.6.1  Dilute the CSE with LRW to get 2λλλ/2, λ/4 (EU/ ml) concentration of Endotoxin or any suitable dilution step.
    5.6.2  Add 100 µL of each Endotoxin concentration to respective tube marked as 2λλλ/2, λ/4. Prepare four sets for each concentration.
    5.6.3  Add 100 µL of Lysate to each tube.
    5.6.4  For negative control add 100 µL of LRW to test tube and add 100 µL of Lysate.
    5.6.5  Vortex gently each tube and insert in triobloc maintained at a temperature 37oC ± 1°C and incubate for 60 min ± 2 min.
    5.6.6  After completion of incubation remove the tubes from triobloc and observe the tubes by inverting gently at 180° for positive or negative results. 
    5.6.7  A firm gel clot that maintains its integrity upon inversion at 180° indicates positive results. 
    5.6.8  Tubes, which indicate no firm gel or show a visible increase in turbidity or viscosity, are reported as negative.
    5.6.9 Note down the results. Find out the endpoint, the lowest concentration that gives positive results is the endpoint. Find out the log endpoint.
    5.6.10  Find out the Geometric mean as follows,
    Geometric Mean  =     Antilog  Ʃ
                                                        f
    Ʃ = sum of log endpoints   f = Number of replicates
    5.6.11  Label claim is verified ok when, Geometric Mean of the endpoint confirms between 0.5λ and 2λ of the label claimed the sensitivity of lysate. Record the results.

    5.7  Test Procedure
    Note: In case of any interference, the sample preparation shall be carried out by pH adjustment using acid, base or suitable buffers, using dispersing agents in case of viscous products or using the reagents of higher sensitivity.
    5.7.1  Sample shall be diluted / prepared as per the requirement and the same sample shall be taken for further testing.
    5.7.2  50 : 50 method
    Details
    Test Solution
        LRW
         C.S.E
    Lysate
    Negative LRW control (NC)
    100µL
    100µL
    Negative product control (NPC)
    50µL
    50µL
    100µL
    Positive Product control (PPC)
    50µL
    50µL (4λ)
    100µL
    Positive LRW control (PC)
    50µL
    50µL (4λ)
    100µL
    NA: Not applicable                µL: Micro litre                 LRW:   LAL reagent water
    5.7.3  Hot spike method
    Details
    Test Solution
    LRW
    C.S.E
    Lysate
    Negative LRW control (NC)
    100µL
    100µL
    Negative product control (NPC)
    100µL
    100µL
    Positive Product control (PPC)
    100µL
    10µL (20λ)
    100µL
    Positive LRW control (PC)
    100µL
    10µL (20λ)
    100µL
    5.7.4  Like the above mentioned in point number 5.7.1 & 5.7.2 reagents shall be added or otherwise, any suitable combinations can be added.
    5.7.5  Gently shake each tube before putting it into triobloc.
    5.7.6  Insert tubes in triobloc maintained at 37° C ± 1° C incubate for a period of 60 ± 2 minutes.
    5.7.7  After completion of incubation remove the tubes from triobloc and observe the tubes by inverting gently at 180° for results. 
    5.7.8  A firm gel clot that maintains its integrity upon inversion at 180° indicates positive results. 
    5.7.9  Tubes, which indicate no firm gel or show visible increase in turbidity or viscosity, are reported as negative.
    5.7.10  Record the observations.
    5.8  Validation
    5.8.1  Validation should be done for 3 batches of each product as per validation protocol. Revalidation shall be carried out as per the validation protocol.
    5.8.2  Validation is satisfactory if both series endotoxin / product and endotoxin/LRW conforms not less than 0.5λ and not more than 2λ of the label claimed sensitivity of lysate. (As per the validation protocol)

    5.9  Interpretation of Results
    5.9.1  The product complies with LAL test if negative results are found in both tubes of NPC and positive results are found in both tubes of PPC.  Positive control (PC) should show positive results in both tubes and negative control (NC) should show negative results in both tubes.
    5.10 Instrument usage details shall be recorded in SOP
    6.0 Abbreviations
    6.1 SOP - Standard Operating Procedure
    6.2 EU - Endotoxin Unit
    6.3 EL - Endotoxin Limit
    6.4 NC - Negative Control
    6.5 MVD - Maximum Valid Dilution
    6.6 MVC - Maximum Valid Concentration
    6.7 ELC - Endotoxin Limit Concentration
    6.8 LRW - LAL Reagent Water
    6.9 CSE - Controlled Standard Endotoxin
    6.10 GM - Geometric Mean
    6.11 NPC - Negative Product Control
    6.12 PPC - Positive Product Control
    6.13 PC - Positive Control
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App