    Calibration of UV / Visible Spectrophotometer

    Calibration of the UV spectrophotometer including control of absorbance using potassium dichromate solution, resolution power using toluene in hexane, limit of stray light and wavelength accuracy.
    Ensure that the connections of the spectrometer are proper.

    Control of Absorbance

    Potassium Dichromate Solution
    Dry a quantity of potassium dichromate by heating to constant weight at 130°C. Weigh & transfer accurately a quantity not less than 57.0 mg & not more than 63.0 mg to 1000 ml volumetric flask. Dissolve & dilute in sufficient 0.005M H2SO4 to produce 1000 ml.
    Measure the absorbance of potassium dichromate solution at the wavelengths given below.
    Calculate the value of A (1% 1cm) for each wavelength.
    A (1% 1cm) = Absorbance X 10000 / Weight of Potassium dichromate in mg.

    Acceptance criteria

    Wavelengths (nm)
    A [1% 1cm]
    Limit
    235.0
    124.5
    122.9 - 126.2
    257.0
    144.0
    142.8 - 145.7
    313.0
    48.6
    47.0 - 50.3
    350.0
    106.6
    105.6 - 108.2
    Stringent limits are adopted from IP & BP for acceptance criteria

    Resolution power

    Record the spectrum of a 0.02% v/v solution of toluene in hexane in the range of 260 nm to 420 nm (before use check the hexane for transmittance, using water as a blank between 260nm to 420nm & use only if transmittance is not less than 97%).

    Acceptance criteria

    The ratio of the absorbance at the maximum at about 269nm to that at the minimum at about 266nm is not less than 1.5.

    Limit of Stray Light

    Prepare a 1.2 % w/v solution of potassium chloride in water.
    Measure absorbance of the above solution at 198.0, 199.0, 200.0, 201.0, 202.0 nm using water as blank.

    Acceptance criteria

    Absorbance is greater than 2.

    Wavelength accuracy, Resolution & Baseline flatness (inbuilt test)

    Attach printer directly to the instrument (instead of the computer).
    Go to รž MODE press F3 key i.e. Maintenance.
    Press ‘1’.
    Press Start/Stop key.
    After screen changes, ensure that nothing is kept in the optical path & press the Start/Stop key again.
    Prints will come after all the three tests are over.

    Acceptance criteria

    For wavelength accuracy: at 656.1 ± 0.3 nm & at 486.0 ± 0.3 nm.
    Resolution: 1.0 nm or less.
    Baseline flatness: ± 0.002Abs.
    Frequency: Once a month.

    Maintenance/ Repairs

    When the instrument does not comply with the requirement/ tolerance range specified above, the instrument should be labeled “OUT OF CALIBRATION” and should be repaired/ serviced.
    After repair/ maintenance, calibrate the instrument.

    REFERENCE

    Indian Pharmacopoeia
    British Pharmacopoeia
    Instruction manual of UV-1700.

    CALIBRATION REPORT OF UV / VISIBLE SPECTROPHOTOMETER

    Page 1 of 2
    SOP Reference –
    Date -
    Instrument No. 
    Frequency
    : Once in a month
    Make –

                 Model –

    1.
    Control of Absorbance:
    Potassium dichromate solution :
    Gross Wt. :  ………………….                                 Balance :
    Tare Wt. :  ………………….
    Net Wt.    : ………………….                                        ………..           ml with 0.005M sulphuric acid.
    Dissolve and dilute 57.0 to 63.0 mg of potassium dichromate in sufficient 0.005M sulphuric acid to produce 1000.0 ml.
    Wavelength (nm)
    Absorbance
    A (1%, 1cm)
    Tolerance
    235.0


    122.9 to 126.2
    257.0


    142.8 to 145.7
    313.0
    47.0 to 50.3
    350.0
    105.6 to 108.2
    Calculation :
                 1%                    Absorbance  x 10000
              A1 cm    = -------------------------------------------------------
                              Weight of potassium dichromate in mg
    Remark : Satisfactory / Not satisfactory.
    Performed by :
    Date :
    Checked by :
    Date :
    2.
    Resolution power :
    0.02% v/v toluene in hexane : 
                 ……….ml   → 100.0 ml with hexane  / ………  ml   → 100.0 ml with hexane
    Absorbance at 269 nm = ……………
    Absorbance at 266 nm = ……………
                              Absorbance at 269 nm
                    Ratio = ---------------------------------- =    ---------------- =
                              Absorbance at 266 nm                             
      (Limit : Not less than 1.5)
    Remark : Satisfactory / Not satisfactory.
    Performed by :                                                                 Checked by:
    Date :                                                                                Date           : 

     

    CALIBRATION REPORT OF UV / VISIBLE SPECTROPHOTOMETER

    Page 2 of 2
    3.
    Limit of stray light:
    Potassium chloride solution
    Gross Wt.                :  ………………….                                  Balance :                 
    Tare Wt. :  ………………….
    Net Wt.    : ………………….                          ……………. ml with water.
    Dissolve and dilute accurately 1.2 g potassium chloride  in sufficient water to produce 100.0 ml.
    Sr. No.
    Wavelength (nm)
    Absorbance
    Tolerance
    1
    198.0
    Absorbance is greater than 2
    2
    199.0
    3
    200.0
    4
    201.0
    5
    202.0
    Remark : Satisfactory / Not satisfactory.
    Performed by :
    Date :
    Checked by :
    Date :
    4.
     Wavelength accuracy (In built test)
    Sr. No.
    Wavelength Standard (nm)
    Wavelength Observed (nm)
    Difference
    Limit (nm)
    1
    656.1
    + 0.3
    2
    486.0
    + 0.3
    Remark : Satisfactory / Not satisfactory.
    Performed by :                                                                 Checked by:
    Date :                                                                                Date           : 
    5.
    Resolution (In built test)
    Sr. No.
    Limit
    Observation
    1
    1.0 nm or less
    nm
    Remark : Satisfactory / Not satisfactory.
    Performed by :                                                                 Checked by:
    Date :                                                                                Date           : 
    6.
    Baseline flatness (In built test)
    Sr. No.
    Limit
    Observation
    1
    ± 0.002 Absorbance
    Remark : Satisfactory / Not satisfactory.
    Performed by :                                                                 Checked by:
    Date :                                                                                Date           : 
    Conclusion : Instrument working is Satisfactory / Not satisfactory.
    Next calibration due on :

    Performed by :
    Date :
    Checked by :
    Date :

    1 comment: Post Yours! Read Comment Policy ▼

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App