    Calibration of Automatic Potentiometric Titrator

    Learn how to calibrate the Automatic Potentiometric Titrator using Disodium tartarate as standard.
    Turn the instrument & printer on.

    Repeatability of Factor Determination

    Go to titer mode.
    Use Disodium tartrate as standard.
    Take all precautions to avoid variations due to weighing & transfer of standard.
    Enter sample weight.
    Find out the factor.
    Take five readings.
    Calculate %RSD & average of five.
    %RSD of five readings should be less than 0.5.

    Calibration of Burette

    Clean the ICB (Interchangeable burette).
    Fill water in the reservoir.
    Fill the burette.
    Go to AUX functions & set Eject volume to 0.10ml.
    Quit & then press Eject.
    Collect the volume in clean, dry, & previously weighed beaker.
    Weigh the beaker & find out the weight of ejected volume, further find out the volume.
    Volume = Weight in g / 0.99602
    Volume should be between 0.0990ml to 0.1010 ml.
    Fill the burette.
    Go to AUX functions & set Eject volume to 1.0ml.
    Quit & then press Eject.
    Collect the volume in clean, dry, & previously weighed beaker.
    Weigh the beaker & find out the weight of ejected volume, further find out the volume.
    Volume = Weight in g / 0.99602
    Volume should be between 0.99ml to 1.01 ml.
    Fill the burette.
    Go to AUX functions & set Eject volume to 10.0ml.
    Quit & then press Eject.
    Collect the volume in clean, dry, & previously weighed beaker.
    Weigh the beaker & find out the weight of ejected volume, further find out the volume.
    Volume = Weight in g / 0.99602
    Volume should be between 9.98ml to 10.02 ml.
    Fill the burette.
    Go to AUX functions & set Eject volume to 20.0ml.
    Quit & then press Eject.
    Collect the volume in a clean, dry, & pre-weighed beaker.
    Weigh the beaker & find out the weight of ejected volume, further find out the volume.
    Volume = Weight in g / 0.99602
    Volume should be between 19.97ml to 20.03 ml.

    Precautions

    Clean the ICB with acetone after calibration of burette is over.

    Frequency

    Once in three months.

    REFERENCE

    User’s guide for Automatic Potentiometric Titrator
    Indian Pharmacopoeia
    A]  Repeatability of factor determination

    Sr. No.

                 Burette Reading (in ml.)

    Factor
    Average
    % RSD
    1
    2
    3
    4
    5
    ACCEPTANCE CRITERIA
    % RSD Not more than 0.5%.

    B]  Calibration of Burette.

        S. No.

    Volume (ml.)

    I

    II

    III

    IV
    Limit (ml.)

    Empty Beaker Wt. (g.)

    Wt. of Beaker + Water (g.)

    II –I

    III / 0.99602

    1

    0.1

     

    0.0990 to 0.1010

    2

    1.0

     

    0.99 to 1.01

    3

    10.0

    9.98 to 10.02

    4

    20.0

    19.97 to 20.03
    CONCLUSION –
    Calibration Complies / Does not comply as per Standard Calibration Procedure.
    Next Calibration Due On -
    Done By :
    Date :
    Checked By :
    Date :

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App