Calibration of Micropipettes

Learn how to calibrate the Micropipettes using distilled water.

1.0 Equipment Required

Micropipettes
Micro tips
Calibrated balance

2.0 Material Required

Distilled water

3.0 Procedure

3.1 Calibrate the micropipettes in a vibration-free room at a constant temperature of water between 20 to 25°C.
3.2 Calibrate the Micropipettes for a volume of 100 and 1000 ml, and for 20 and 200 ml.
3.3 Set the desired volume using the volume-setting knob.
3.4 Carefully fit tip into tip cone and flush with distilled water by pipetting the adjusted volume at least 3 – 4 times.
3.5 Carefully aspirate the liquid and pipette the sucked volume of distilled water into a tare 10 ml volumetric flask.
3.6 Repeat at least ten times and record the result.
3.7 Calculate the pipetted volume, taking the temperature into account.
3.8 After calibration, immediately put the “CALIBRATION” tag duly filled with necessary information

4.0 Precaution

4.1 Use an analytical balance with a readability of 0.1 mg.

5.0 Frequencies

Quarterly

6.0 Limits/ Acceptance Criteria

Micropipettes ID
Volume in ml
Tolerance Limit in ml
Min
Max
A
100
99.9
100.1
1000
999.0
1001.0
B
20
19.98
20.02
200
199.8
200.2

7.0 Abbreviation

SOP: Standard Operating Procedure





is a prominent Pharmaceutical Quality Assurance expert, consultant and the founder of Pharmaguideline. With over 22 years of hands-on experience in cGMP-compliant manufacturing environments, he specializes in establishing validation protocols, sterile area controls and data integrity systems. Ankur routinely interprets international regulatory frameworks (including FDA, EMA and ICH guidelines) to help global pharmaceutical professionals ensure strict regulatory compliance and operational excellence. Connect with Ankur on LinkedIn. Need Help: Ask Question

No comments:

Post a Comment

Please don't spam. Comments having links would not be published.