    Calibration of Dissolution Testing Apparatus

    Learn the calibration of Dissolution Test Apparatus using Disintegrating Prednisone tablets.

    RPM  CALIBRATION

    1.  Enter the required RPM value in the range of 30 to 200.
    2.  Press ENTER key to terminate. The SET RPM key LED will now be put off. If the RPM value is not in the above-mentioned range error message Er is displayed on the RPM display.
    3.  If the error message is displayed then go back to resetting. If the error message "Er" is displayed on the RPM display and the ENTER key is pressed then the last correctly set value of RPM for that step will be accepted and the SET RPM LED will be put off.
    4.  Suppose RPM value 100 is to be entered. First press the SET-RPM key. The SET-RPM key LED will be illuminated., Next press number 1 key once following by 0 key twice, Then press ENTER key. The SET-RPM key LED will go OFF, Then the RPM DISPLAY will show 100.
    5.  Record the RPM manually & enter in the calibration log.

    TEMPERATURE CALIBRATION

    1.  Press the SET key. The SET TEMP key LED will be illuminated TEMP.
    2.  Enter the required temperature value in the range of 30 °C to 100 °C.
    3.  Press ENTER key to terminate. The SET TEMP key LED will now be put off. If the Temp. is not in the above-mentioned range then the error message Er is displayed on the TEMP display. If the error message is displayed then go back to resetting.
    4.  If the error message "Er" is displayed on the TEMP display and the ENTER key is pressed then the last correctly set value of TEMP for that step will be accepted and the SET TEMP LED will be put off.
    5.  Set the temp. 35°, 37° & 40°C and check the temperature of the bowl with the thermometer & record the data in calibration log.
    6.  To view the set temperature presses the SET TEMP key again. Then the TEMP DISPLAY will show set temp in °C.
    7.  If the calibration is not found within the specified limits immediately inform to QC head.  
    Affix the calibration label on the instrument.

    CALIBRATION FOR DISINTEGRATING TYPE PREDNISONE TABLETS 50 MG

    1.  Ensure that mains of instrument and supply are “ON”.
    2.  Set the time program for 30 minutes, temperature 37° C + 0.5° C and for 50 RPM.
    Fit the paddle apparatus.
    3.  Place 900 ml of deaerated water.
    4.  After achieving the temperature place one tablet in each bowl and start the time program.
    5.  After the completion of time program withdraw the sample solution from each bowl and filter through the whatman filter no. 41 in individual tubes. Discard first 50 ml filtrate, take 5 ml of filtrate and dilute with water to 25 ml and take the absorbance at 242 nm against the water.
    6.  Prepare the standard solution by dissolving the 25 mg of prednisone to 50 ml of water, further dilute it 5 ml to 50 ml with the water, further dilute 5 ml to 25 ml with water and take the absorbance at 242 nm against the water. (If the standard is not dissolved first add not more than 2.5 ml of alcohol)
    7.  Calculate the content of prednisone. Also, carry out above procedure for 100 rpm. Enter the data in the calibration log.
    8.  Also, carry out above procedure by changing the basket apparatus for 100 RPM and enter the data in calibration log.
    •  If the calibration is not found within the specified limits immediately inform to QC head.
    •  Affix the calibration label on the instrument.
    •  In the case of noticing any malfunctioning of the instrument, inform the maintenance dept. and Head of the dept.

    Related: Why Dissolution Test Apparatus Calibration with Salicylic Acid Tablets was Stopped?

    Checking of physical parameters for the apparatus:

    Basket

    Make Identity of the each basket which is in use and make the inventory. Check the physical parameters for the each basket like appearance, opening, height, total height of basket, ID of basket, external diameter of basket, external diameter of ring, vent hole diameter, height of coupling disk. All parameters should fall within the limit as given in the calibration log.

    Paddle

    Make identity of the each paddle which is in use and make the inventory. Check the physical parameters for the each paddle like appearance, height, shaft diameter, blade upper chord, lower chord, height, radius (disk), thickness, and distance from bottom, distance shaft axis and vertical axis of vessel. All parameters should be fall within the limit as given in the calibration log.

    Checking of wobbling, Rotations, and distance parameters for the apparatus

    Wobbling

    •  Fix the paddle/basket apparatus to the instrument, lower the apparatus.
    •  Fix the wobbling meter on the top of the bowl.
    •  The needle of wobbling meter should touch to the surface of the apparatus.
    •  Measure the wobbling on the meter.

    Rotation

    •  Fix the paddle/ basket apparatus to the instrument.
    •  Set the RPM 50/100.
    •  The needle of tachometer should touch to the surface of the apparatus.
    •  Press the “start” key of tachometer to start
    •  Fix the tachometer with apparatus.
    •  Start the rotation.
    •  The tachometer will show you result on display.

    Distance from the apparatus

    •  Fix the paddle/basket apparatus to the instrument, lower the apparatus.
    •  Insert the vernier calipers between the bottom of the apparatus and the bottom of the bowl.
    •  Measure the reading on the display.

    Related: Tablet Dissolution Test in Different Stages (S1, S2 and S3)

    TEMP./RPM CALIBRATION

    Model No.                   :  
    Make                   : 
    Instrument Code No.    :
    Thermometer code #      :

    Sr  #
    Type of calibration
    Set value
    Observed value
    Limits
    1
    2
    3
    4
    5
    6
    1
    35° C
    34.5 - 35.5°C
    Temp °C
    37° C
    36.5 - 37.5°C
    40° C
    39.5 - 40.5°C
    2
    50
    48-52
    RPM
    100
    96-104
    150
    144-156
    Calibration Status         :  Satisfactory / Not Satisfactory
    Next Calibration Due    :
    Calibrated By   :                                                                             Checked By     : 
    Date                  :                                                                            Date                :


    CALIBRATION LOG (FOR DISINTERATING TYPE)

    Model                          :  
    Make                   : 
    Instrument Code No.    :
    Lot no of  tablet  :
    Std. Preparation : (Code of WRS___________)
    Transferred __________ mg (25 mg ) to 50 ml volumetric flask, dissolved and diluted to 50 ml with water. 
    Diluted _________( 5 ml ) to 50 ml with water and mix. Diluted _________( 5 ml ) to 25 ml with water and mix.

    Weight of tablet
    Dissolution time
    Diluted to
    (5 ml to 50 ml)
    Diluted to
    (5 ml to 25 ml)
    UV reading at 242 nm)
    Remark
    From
    To
           1.
           2.
           3.
           4.
           5.
           6.

    Sr. No.
    Set apparatus
    Observed
     value
    50 rpm
    100 rpm
    Limits
    1.
    Apparatus I
    Min.:
    Max.:
    Mean.:
    2.
    Apparatus II
    Min.:
    Max.:
    Mean.:

    Calibration Status         :  Satisfactory / Not Satisfactory
    Next Calibration Due   :
    Calibrated By   :                                                                                   Checked By     :
    Date               :                                                                                     Date                : 

    PHYSICAL PARAMETER (PADDLE)

    Model No.                   :  
        Make              : 
    Instrument Code No.    :

    Sr.
    No.
    Parameter
    Specifications
    Observations
    Remark
    1
    2
    3
    4
    5
    6
    mean
    1.
    Shaft diameter 
    9.4 - 10.1 mm
    2.
    Blade upper chord
    74.5 ± 0.5 mm
    3.
    Lower chord
    42.0 + 1.0 mm
    4.
    Paddle Height
    19.0 + 0.5 mm
    5.
    Thickness
    4.0 + 1.0 mm
    6.
    Distance from bottom
    25 +2 mm.
    Calibration Status        :  Satisfactory / Not Satisfactory
    Next Calibration Due    :
    Calibrated By   :                                                                    Checked By     :
    Date                 :                                                                    Date                 : 

    PHYSICAL PARAMETER (BASKET)

    Model No.                   :  
    Make                   : 
    Instrument Code No.    :

    Sr.
    No.
    Parameter
    Specifications
    Observations
    Remark
    1
    2
    3
    4
    5
    6
    mean
    1.
    Total height of basket
    37.0 +3.0 mm
    2.
    Height of screen
    27.0 ± 1.0 mm
    3.
    ID of screen
    20.2 + 1.0 mm
    4.
    External diameter screen
    22.2 + 1.0 mm
    5.
    External diameter of ring
    25.0 + 0.02 mm
    6.
    Hole diameter
    2.0 +0.5 mm
    7.
    Height of coupling disk
    5.1 + 0.5 mm
    8.
    Distance from bottom
    25 +2 mm.
    Calibration Status         :  Satisfactory / Not Satisfactory
    Next Calibration Due    :
    Calibrated By   :                                                                                   Checked By     :
    Date                 :                                                                                   Date                 :

    CHECKING OF WOBBLING, ROTATION AND DISTANCE

    Model
    Code No.

    Test unit
    Paddle wobble
    Basket wobble
    Paddle rotation
    D1 mm
    D2 mm

    USP standards
    (2.0 mm)
    (2.0 mm)
    (4 % of set rpm)
    (25.0+ 2.0 mm)
    (25.0+ 2.0 mm)

    50 RPM
    100 RPM

    1.

    2.

    3.

    4.

    5.

    6.

    D1
    Distance between the bottom edges of the blade to the lowest inner surface of the vessel.

    D2
    Distance between the bottom edges of the basket to the lowest inner surface of the vessel.


    Calibration Status         :  Satisfactory / Not Satisfactory
    Next Calibration Due    :
    Calibrated By   :                                                                       Checked By     :
    Date                 :                                                                       Date                 : 

    5 comments: Post Yours! Read Comment Policy ▼

    1. Good one sir! what is the claim of tablet(Prednisone) through which we calibrate the dissolution test apparatus.....

      ReplyDelete
    2. Sir what is the calibration frequency?

      ReplyDelete
    3. In dissolution calibration prednisone tablet % drug release is higher in basket than paddle why?

      ReplyDelete
    4. What is the FDA acceptance of dissolution apparatus calibration using IP reference standard of Prednisone tablets. Please advise

      ReplyDelete

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App