Calibration of Brookfield Viscometer | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    Calibration of Brookfield Viscometer

    Learn how to verify the performance of Brookfield Viscometer using 5.00 cps, 50.00 cps and 5000.00 cps standards.

    PROCEDURE

    • Ensure that all the connections of the instrument are proper.
    • Turn the viscometer ON.
    • Take out the viscosity standard solutions in 500 ml beaker, adjust the temperature to 250C. Attach the required spindle, adjust required speed & measure the viscosity, note down the constant reading.
    • Viscosity standards, speeds, spindles required & tolerance limits are as follows:
    Sr.
    No.
    Viscosity of standard in cps
    Actual viscosity of standard in cps
    Spindle No.
    Entry Code
       RPM
    Tolerance (in cps)
    Minimum
    Maximum
     1.
     5.00
    4.70
     ULA
    00
    60
    4.55
    4.85
     2.
     50.00
     48.40
     LV1
    61
    60
    46.9
    49.9
     3.
    5000.00
    4820.00
    LV3
    63
     10
    4651.83
    4988.17

    Acceptance criteria

    • Viscosity should be within the tolerance limits.
    • Do not transfer the viscosity solution to the original container back after use.

    Precautions

    • Take due care to avoid any contamination of the standard solutions.
    • Avoid air bubble entrapment during transfer of standard solutions.

    Frequency

    Once in 6 months.

    QUALITY ASSURANCE DEPARTMENT
    CALIBRATION REPORT OF BROOKFIELD VISCOMETER
    Reference – 
    Date -
    Frequency - Once in 6 Months
    Instrument No. -
    Make – Brookfield Engineering Laboratories, Inc.
    Model – LVDV-II + Pro
    Take out the viscosity standard solutions in 500 ml beaker, adjust the temperature to 250C & attach the required spindle, adjust required speed & measure the viscosity, note down the constant reading.

    Sr. No.
    Viscosity Std. in CPS
    Actual Viscosity of Std. in CPS
    Observed Viscosity in CPS
    Tolerance (in CPS)
    Remark
    (S / NS)

    Minimum
    Maximum

    1
    5.00
    4.70
    4.55
    4.85

    2
    50.00
    48.40
    46.9
    49.9

    3
    5000.00
    4820.00
    4651.83
    4988.17

    Conclusion : Calibration is satisfactory (S) / not satisfactory (NS).

    Next Calibration Due On :

    Done By :
    Date :
    Checked By :
    Date :

    1 comment: Post Yours! Read Comment Policy ▼

    1. How to i have know the spindle,Rpm for standard fluid in Brookfield odometer LVII+.

      ReplyDelete

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App