    SOP for Calibration of Colony Counter

    Standard operating procedure of evaluation of performance of digital colony counter.

    1.0 OBJECTIVE

    To describe the procedure to calibration of colony counter.

    2.0 SCOPE

    This SOP is applicable to calibration of colony counter.

    3.0 RESPONSIBILITY

    Microbiologist - Quality Control

    4.0 ACCOUNTABILITY

    Manager – Quality Control

    5.0 PROCEDURE

    5.1 Write counting up to 100 on a white paper.
    5.2 Operate the equipment as per SOP of operation of colony counter.
    5.3 Press the surface of counter screen 100 times with pointer/marker and then check the number displayed on the screen.
    5.4 Reset to 000 and again press the manual button 100 times and check the number displayed on the screen. In both cases, the number displayed on the screen is also 100.
    5.5 Repeat this procedure for 5 times to ensure that the calibration procedure will give a consistent result.
    5.6 If the equipment does not produce required calibration results or its response is poor then it should be labeled “Out of Calibration” and should be repaired.
    5.7 After completion of calibration procedure record in the calibration record of colony counter Annexure I.
    5.8 Frequency: The calibration frequency of digital colony counter is once in a month.

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure

    ANNEXURE- I 
    CALIBRATION RECORD OF COLONY COUNTER
    Asset ID  No.
    :

    Ref.  SOP No.
    :

    Make
    :

    Model No.
    :

    Effective Date         :                                                                       
    Location                  :
    Date of Calibration  :
    Next Due Date         :
    Surface pressed
    Number displayed on screen
    Manual button pressed
    Number displayed on screen
    100 times

    100 times

    100 times

    100 times

    100 times

    100 times

    100 times

    100 times

    100 times

    100 times

    Observation: The Colony Counter is working satisfactorily /not working satisfactorily, requires/ does not require maintenance.

    Calibrated By:                                                                                      Reviewed By:
    Date:                                                                                                    Date:
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App