SOP for Calibration of Hot Air Oven | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    SOP for Calibration of Hot Air Oven

    Standard operating procedure of evaluation of hot air oven used in Quality Control.

    1.0 OBJECTIVE

    To describe the procedure for the calibration of the hot air oven

    2.0 SCOPE

    This SOP is applicable for the calibration of hot air oven.

    3.0 RESPONSIBILITY

    Officer /Executive - Quality Control

    4.0 ACCOUNTABILITY

    Manager - Quality Control.

    5.0 PROCEDURE

    5.1 For the calibration of hot air oven use a standard thermometer ranging up to 300º C.
    5.2 Start the calibration procedure after 1 hour of staring the oven.
    5.3 Set the oven at desired temperature.
    5.4 Put the standard thermometer for 30 minutes on the upper shelf of the oven and close the door of the oven.
    5.5 After 30 minutes open the door of the oven and read the temperature of standard thermometer randomly and match the observed temperature of the thermometer.
    5.6 Repeat the above procedure by putting the thermometer in lower shelf for 30 minutes.
    5.7 Record the observed temperature in calibration record of hot air oven as per Annexure no. I
    5.8 The observed temperature of the thermometer in both shelves should be ± 2.0º C tolerance limit to the set temperature value.
    5.9 Frequency: Once in a month.
    Related: SOP for Validation of Oven for Depyrogenation

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure

    ANNEXURE- I
    CALIBRATION RECORD OF HOT AIR OVEN
    Inst. ID  No.
    :

    Ref.  SOP No.
    :

    Make
    :

    Model No.
    :
      
    Date of Calibration     :
    Next Due Date           :
    Location                     :
    Frequency: Monthly  :
    Location
    Set Temperature
    Observed Temperature 
    Tolerance
    Observed By
    Upper shelf


    ±2.0º C

    Lower shelf


    ±2.0º C

    Remark: The calibration procedure of instrument are within tolerance limit /out of tolerance limit and gives satisfactory / not satisfactory results.
    Analyst                                                                                                     Checked by
    Date                                                                                                          Date
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App