    SOP for Calibration BOD / Bacteriological Incubator

    Standard operating procedure to evaluation of performance of incubators used to incubate the microbial cultures.

    1.0 OBJECTIVE

    To describe the procedure for the calibration procedure of BOD/bacteriological incubator.

    2.0 SCOPE

    This SOP is applicable for the calibration procedure of BOD/ bacteriological incubator.

    3.0 RESPONSIBILITY

    Microbiologist - Quality Control

    4.0 ACCOUNTABILITY

    Manager - Quality Control

    5.0 PROCEDURE

    5.1 Ensure that material under test not to be disturbed during the calibration procedure.
    5.2 Calibration of the incubator is carried out by Five Pt 100 probes situated on four corners and one in center of the chamber inside the incubator.
    5.3 For calibration of incubator scan the all four Pt 100 probes (channels) by pressing the SCANER key on the universal controller of incubator for 24 hrs.
    5.4 After scanning take the print out of temperature mapping of 24 hrs duration by pressing the PRINT key on the universal controller of the incubator.
    5.5 Record the minimum and maximum temperature attained during 24 hrs temperature mapping and record them in calibration record of the bacteriological incubator as per Annexure I.
    5.6 Observe that all temperature values are within ±2.0º C tolerance limit to the set temperature value.
    5.7 Frequency: Yearly once

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure

    ANNEXURE- I
    Calibration record of BOD/Bacteriological Incubator
    Asset ID  No.
    :

    Ref.  SOP No.
    :

    Make
    :

    Model No.
    :

    Date of Calibration :
    Next Due Date        :
    Frequency               : Yearly
    Probe No.
    Set Temperature
    Observed Temperature 
    Tolerance
    Observed By
    Min. Temp.
    Max.Temp.




    ±2.0º C





    ±2.0º C





    ±2.0º C





    ±2.0º C





    ±2.0º C

    Remark: The calibration procedure of instrument are within tolerance limit /out of tolerance limit and gives satisfactory / not satisfactory results.

    Calibrated By:                                                                                       Reviewed By:
    Date:                                                                                                     Date:
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App