    SOP for Particle Counter

    Standard operating procedure of Particle Counter used to count air born particles in clean area.

    1.0 OBJECTIVE

    To lay down a procedure for operation of the particle counter.

    2.0 SCOPE

    The procedure is applicable for operation of particle counter “Met one 3400”.

    3.0 RESPONSIBILITY

    Officer/ Executive - Quality Control

    4.0 ACCOUNTABILITY

    Manager - Quality Control

    5.0 PROCEDURE

    5.1 Logging in: Switch ON the main power, enter ‘Counter Navigation’ then enter ‘Factory’.
    5.1.1 There are two methods to login, depending on whether the 3400 has first been turned on or if another user is already logged in.
    • To log in after the Model 3400 is initialized, touch the screen. When password protection is enabled, users must log in to proceed.
    • To login to the unit with another user already logged in, touch the Lock icon to log out the previous user, then touch the Lock icon again to login.
    5.1.2 A login screen appears like the one. Touch the Name field to enter the username and a keyboard entry screen appears.
    5.1.3 Enter the username by touching the screen.
    5.1.4 When the name is entered, touch the Enter button on the screen.
    5.1.5 Touch the Password field to invoke the keyboard screen and enter the password via the touchscreen.
    5.1.6 Press Enter
    5.1.7 If login is successful, the Main screen should appear, similar to the one. If the login is unsuccessful, a Login Failed message appears and the password should be re-entered.

    5.2. Taking counts

    Always remove the protective cap from the inlet tube before performing counts or damage to the counter will occur.
    5.2.1 If still in place, remove the protective cap from the inlet tube (on top of counter).
    5.2.2 Make sure the Main screen display ‘Counter Navigation’. Press ‘Sample’ on the screen.
    5.2.3 To start counting particles, press the Run icon, on the touchscreen.
    Particle counting begins.
    • The Run icon becomes a Stop icon until the count is completed.
    • The display indicates the counts and any other parameters measured.

    Note: Count or environmental alarms will be indicated by red color and flashing text at the bottom of the touchscreen, as well as an audible alarm if one was set as described in “Sounds Tab”.
    5.2.4 The count stops automatically. To manually stop counts, press the stop icon. When the count is complete, the total number of particles counted during the count time displays as well as the values for values for any other parameter measured (relative humidity, temperature, air velocity).
    Refer to adjust the settings, such as alarm limits, units of measurements, and time and date.

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App