    SOP for Calibration of Dry Bath

    Standard operating procedure of calibration of Dry Bath using calibrated thermometer.

    1.0 OBJECTIVE

    To ensure that the dry bath is operated and calibrated properly.

    2.0 SCOPE

    This SOP is applicable for operation and calibration of the dry bath.

    3.0 RESPONSIBILITY

    Microbiologist - Quality Control

    4.0 ACCOUNTABILITY

    Manager- Quality Control

    5.0 PROCEDURE

    5.1 General cleaning procedure

    5.1.1 Ensure that the power supply to the dry bath is switched OFF.
    5.1.2 De-dust the dry bath daily externally with a clean dry cloth.
    5.1.3 Once in a week remove adhered dust by wet mopping using soap solution. Afterward, wipe the surface with a clean dry cloth to remove the moisture.
    5.1.4 Mop the interior surfaces with a clean dry cloth, daily.
    Related: SOP for Water Bath

    5.2 Operating Procedure

    5.2.1 Ensure that the dry bath is properly connected to the power supply.
    5.2.2 Switch ‘ON’ the main switch and then the instrument switch.
    5.2.3 Set the required temperature to 37 °C by pressing the 'SET knob and soft keys.
    5.2.4 Monitor the temperature daily during operation.

    5.3 Calibration Procedure

    5.3.1 Switch ON the instrument.
    5.3.2 Put the glycerol solution in the five-hole of the dry bath. Four corners and one center
    5.3.3 Leave for 15 minutes for stabilization.
    5.3.4 Put the calibrated thermometer for each location one by one
    5.3.5 Record the results after stabilization as per Annexure – I
    5.3.6 Frequency of calibration – Monthly once

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure
    6.2 No. - Number
    6.3 QA - Quality Assurance
    6.4 QC - Quality Control
    6.5 CCF - Change control format

    ANNEXURE - I
    CALIBRATION REPORT OF DRY BATH
    Asset ID  No.
    :
    Ref.  SOP No.
    :
    Make
    :
    Model No.
    :

    Observation:
    Sr. No.
    Standard value
    (in  ºC )
    Set  value
    (in  ºC )
    Tolerance
    Remark
    1.
    37 ºC
    ± 1ºC
    2.
    37 ºC
    ± 1ºC
    3.
    37 ºC
    ± 1ºC
    4.
    37 ºC
    ± 1ºC
    5.
    37 ºC
    ± 1ºC

    Remarks:    The instrument is calibrated as per the SOP and shows satisfactory / does not satisfactory results as per calibration procedure.
    Calibrated on                 : ____________________
    Next calibration due on   : ____________________

    Done by:                                                                                          Checked
    Date:                                                                                                Date:
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App