SOP for Cleaning of Atomic Absorption Spectrophotometer : Pharmaceutical Guidelines

SOP for Cleaning of Atomic Absorption Spectrophotometer

Standard operating procedure to clean the atomic absorption spectrophotometer.

1.0 OBJECTIVE

To clearly lay down procedure for cleaning.

2.0 SCOPE

This procedure is applicable to Atomic absorption spectrophotometer installed in quality control laboratory.

3.0 RESPONSIBILITY

3.1 Doing : Technical Assistant /Executive
3.2 Checking : Executive/Manager

4.0 ACCOUNTABILITY

Head of the Department

5.0 PROCEDURE

5.1 After completion of analysis, do the following.
5.2 Remove the used glassware.
5.3 Aspirate with blank solution for one minute to remove any residue remained in the suction pipe.
5.4 Clean the suction pipe with distilled water for four times.
5.5 Switch ‘OFF’ the instrument.
5.6 Switch ‘OFF’ the mains.
5.7 Clean the burner slit using wooden strip (Provided by manufacturer specially for this purpose) or using hard paper.
5.8 Remove the lamp from instrument and keep it in drawer.
5.9 Clean the outer surface of instrument with dry clean cotton cloth and remove any dirt or spots by mopping with isopropyl alcohol and dry it with tissue paper. Record in the format of instrument log card.

6.0 ABBREVIATION 

6.1 SOP - Standard Operating Procedure





Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of Pharmaceutical Guidelines, a widely-read pharmaceutical blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
[email protected] :liamENeed Help: Ask Question


Click Here

No comments: Read Comment Policy ▼

Post a Comment

CURRENT JOBS

Show All ❭❭Jobs by PharmaJobs

Write a Powerful CV

Click Here


Advertise Here

Get App for News Updates

Scan to Download

Android App
Android App

Recent Posts