Performance Qualification Protocol for Sterilization and Depyrogenating Tunnel | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Aug 7, 2011

    Performance Qualification Protocol for Sterilization and Depyrogenating Tunnel

    Validation Protocol for Sterilization and Depyrogenating Tunnel used in sterile production.

    1.0  PERFORMANCE QUALIFICATION TEST

    Following Performance Qualification studies shall be carried out to ensure the equipment for proper operation and its ability to sterilize and depyrogenate the washed vials at the set parameters, repeatedly & consistently.
    ·  Air Velocity measurement.
    ·  HEPA FILTER Integrity test by DOP / PAO Aerosol test.
    ·  Air flow pattern test.
    ·  Non-viable airborne particle count test.
    ·  Heat Distribution studies.
    ·  Heat Penetration studies.
    ·  Endotoxin Challenge study.

    To qualify the test the equipment should fulfill the acceptance criteria described in the individual test procedures. All instruments shall be calibrated before starting and after completion of validation studies.

    After completion of the qualification tests all the data generated will be compiled to evaluate ability of the sterilization and depyrogenation tunnel to sterilize & depyrogenate different articles at the set operating parameters.

    2.0  TEST CALIBRATION EQUIPMENTS

    S. No
    Equipment Name
    Equipment I.D
    Calibration on
    Calibration Due on
    Verified By
    2.1  QUALIFICATION OF THE EXECUTION TEAM:
    The execution team should be trained for execution of operation Qualification of the equipment & Record should be maintained in Annexure 1
    2.2.1  AIR VELOCITY MEASUREMENT
    2.2.1.1  OBJECTIVE
    To demonstrate that the system is capable of delivering air velocities, as per the Requirement, to maintain continuous laminarity of HEPA Filter installed in tunnel.
    2.2.1.2  EQUIPMENT USED
    Digital Anemometer
    2.2.1.3   METHOD
    A)  This test shall be performed by a trained personal and training record shall be attached with report.
    B)  This test shall be performed at least 30 minute after the tunnel has been Started Heater should be off.
    C)  Measure Air velocity 6 inches below the filter at 5 locations (four corner filter and one center), calculate the average velocity of air coming from supply.                               
    D)  Calculate the velocity of air coming from supply filter should be measured in Feet/min.
    E)  Record the Data in Annexure II.

      V1                                         V2
                          
                              V3


     V                                          V5








    Average velocity (A) = V1+V2+V3+V4+V5/5
                                A  = Average     
    2.2.1.4  ACCEPTANCE CRITERIA
    Average velocity across HEPA Filter should be within the range of 90±20 Feet/min
    Average velocity across HEPA Filter in sterilization zone should be within the range of 150-200Feet/min
    2.2.2  HEPA FILTERS LEAKAGE TEST (DOP/PAO TEST)
    2.2.2.1  OBJECTIVE
    To verify the integrity of HEPA FILTERS installed in the sterilization and depyrogenation tunnel.
    2.2.2.2  TEST APPARATUS                                         
    •  Aerosol Generator
    •  Aerosol photometer
    •  Compressed Air

    2.2.2.3  METHOD
    a)  Filter testing shall be performed only after operational air velocities have been verified and adjust wherever necessary.
    b)  Position the smoke generator and introduce (DOP/PAO) smoke into the stream of the HEPA Filter. The concentration of (DOP/PAO) aerosol is 80-120µg per liter of air.
    c)  Programme the instrument at 100% concentration for upstream.
    d)  Scan the downstream side of the HEPA with an appropriate photometer probe.
    e)  The probe shall scan the entire filter face and frame at a position about 1 inch from the face of the filter.
    f)  Scanning shall be done at the rate of 2 feet/minute.
    g)  Record the observation in Annexure III
    2.2.2.4  ACCEPTANCE CRITERIA
    During scanning percentage of DOP or PAO penetration shown by photometer should not be more than 0.01% of the upstream concentration through the filter media and should be zero through mounting joints.
    2.2.3  AIR FLOW LAMINARITY:
    2.2.3.1  OBJECTIVE
    To determine the air flow pattern of HEPA Filter installed in sterilization and depyrogenation tunnel
    2.2.3.2  EQUIPMENT USED
    Titanium Tetra Chloride (Chemical Used).
     Video Camera. 
    2.2.3.3   METHOD
    TITANIUM TETRA CHLORIDE
    a)  Take the glass rod with cotton or sponge tied to it.
    b)  Dip the rod in Titanium Tetra Chloride solution and place the stick at Downstream of HEPA Filter.
    c)  Check the air flow direction at the Downstream of the filter face.
    d)  Observe & record the same with video camera.
    2.2.3.4  ACCEPTANCE CRITERIA
    The stream of air should be unidirectional and non-turbulent upto working zone.
    2.2.3.5  Record the result in Annexure V

    2.2.4  AIR BORNE NON-VIABLE PARTICLE COUNT TEST
    2.2.4.1  OBJECTIVE
    To establish that at different locations within the tunnel, a count size of particles per cubic meter is within the limit.
    2.2.4.2  EQUIPMENT USED
    Air Borne Particulate Counter, duly calibrated with traceability certificate.
    2.2.4.3  METHOD
    a)  The test should be perform only when the earlier test have been performed (air velocity, filter integrity, air laminarity) and found to be conforming to the requirement. The particle count test should be performed by qualified and/ or trained personnel only.
    b)  Calculate the number of location by the following formula
    No. of sampling location: NL=√A
    Where:
    NL=The minimum number of sampling locations (Rounded up to a whole number)
    A=The area of the HEPA Filter installed in tunnel in m2
    c)  Use calibrated particle analyzer for non viable particle count and consider 0.5µ and 5.0 µ
    d)  Perform 3 consecutive cycles.
    e)  Record the data in Annexure VIII
    2.2.4.4  ACCEPTANCE CRITERIA    
    The system shall capable of achieving the desired cleanliness grade in the subjected critical work locations in the tunnel as per the designed specified limit                   
                     CRITERIA
    ACCEPTANCE

     Non viable Air particle count test
    Maximum number of permitted particles per cubic meter equal to or above
              GRADE
    ≥ 0.5 µm
     ≥5.0 µm
    GRADE A
    3520
    29
    2.2.5  HEAT DISTRIBUTION STUDY
    2.2.5.1  OBJECTIVE
    The objective of the test is to ensure that:
    The Sterilizing & Depyrogenating Tunnel when operated with Empty Chamber is capable of producing the temperature profiles as per the temperature set points set in the PLC of the equipment. The temperature distribution is uniform throughout the sterilization period.
    Three run to be performed to qualify the measurement of the temperature throughout the chamber during the sterilization cycle.
    2.2.5.2  Equipment Used
    Data Logger with Minimum 12 Probes duly calibrated.
    2.2.5.3  METHOD
    Record the set parameters of the sterilization cycle to be operated during the test. Suspend the probe in the chamber in different position in such a way that probes don’t touch any metallic surface. Record the position of the probe in a representative schematic manner.
    Connect the probes to suitable data logger, which can scan and print the actual temperature observed at different locations with respect to time.
    Operate the Sterilizing & Depyrogenation Tunnel Also start the data logger to record the actual temperatures with respect to time.
    After completion of sterilization cycle switch off the data logger.
    Collect print out from the printer of Sterilizing & Depyrogenation Tunnel.
    Download the data from the data logger in the computer for the data analysis and printing enclosed the printout obtained from the data logger.
    2.2.5.4  ACCEPTANCE CRITERIA:
    The Temperature at each Temperature probe should be ≥300°C during the cycle.
    2.2.5.5 Observations and Results In Annexure VI
    Record the temperature at various locations

    2.2.6  HEAT PENETRATION STUDY
    2.2.6.1  OBJECTIVE
    Objective of this test is to ensure that, heat is sufficiently penetrating into the innermost  portion of the Vial subjected for sterilization & Depyrogenation to achieve desired temperature during the sterilization & Depyrogenation cycle.
    Loaded chamber Heat penetration studies must be conducted for three consecutive cycles with temperature probes.
    The recovery of Endotoxin Concentration after exposing to Depyrogenation tunnel should show more than 3 log reduction.
    Three run to be performed to qualify the measurement of the temperature throughout the tunnel by seven (during the sterilization cycle).
    2.2.6.2  EQUIPMENT USED
    Data Logger with 12 Probes duly calibrated.
    2.2.6.3  METHOD
    Conduct the study with loaded chamber for three consecutive cycles with temperature probes and Endotoxin vials.
    Suspend the 12 probes inside the vial and put into tunnel for Heat Penetration Study.
    Record the position of the probes in a representative schematic form.
    Insert 9 Endotoxin Vials (Marked vials) having 10000 EU each along with the temperature sensors.
    Connect the probes to suitable data logger, which can scan and print the actual temperature observed at different locations with respect to time.
    Operate the Sterilizing & Depyrogenating Tunnel  also start the data logger to record the actual temperatures within the Sterilizing & Depyrogenating Tunnel with respect to time.    
    BELT SPEED CAN BE CALCULATED BY USING THE FOLLOWING FORMULA:

          (Vial diameter)2 × cos30o × washing machine output
                              Tunnel conveyor width

     STERILIZATION HOLD TIME:            Length of sterilizing zone
                                                                                Conveyor speed

    2.2.6.4  ACCEPTANCE CRITERIA
    Throughout the dwell time, all temperature measured in the chamber is ≥ 300°C. The recovery of endotoxin concentration after exposing in sterilization and depyrogenation should show at least 3 log reduction.
    3.0  ACCEPTANCE CRITERIA
     S.No.
    TEST  PERFORMED
    ACCEPTANCE CRITERIA
    1.
    Air velocity test
    Average velocity across the HEPA filter should be within the range of 90 ± 20% FPM.
    2.
    HEPA Filter integrity testing
    Leakage in the filter should not be more than 0.01% and should be zero through mounting joints.
    3.
    Air flow pattern
    Air flow pattern should be unidirectional and non turbulent.
    4.
    Non Viable particle counter
    Acceptable limit of
    0.5µ Particles is 3520  and 5.0µ is 29 in 1M3 of air.
    5.
    Heat distribution studies
    Temperature should be uniform throughout the cycle
    6.
    Heat penetration studies
    Should show at least 3 log reduction after exposing
    Sterilization temperature

    4.0  RESULT:

    The results obtained after execution of the protocol, will be presented in the validation report.  All results will be compared against the acceptance criteria. Any deviation failure to meet these specifications will be duly documented. A deviation indicates variance from the acceptance criteria however, does not necessarily mean that the qualification as unacceptable. During execution of this protocol, if any deviation is noticed, the person executing the protocol initiates a deviation report and provides the detail description of the deviation. Investigation may be performed to identify root cause for the deviations or failures and corrective action will be taken to minimize such deviations or failures in the future.

    5.0  CONCLUSION:

    Conclusion will be drawn after compilation and evaluation of result.

    ANNEXURE 1
    Training for Performance Qualification of sterilization and depyrogenating Tunnel
    Name of the trainer  :       
    Designation              :
    Date of training        :
    Duration                  :
    Sr. No.
    Name of Trainee
    Department
    Signature

    Comments  _____________________________________________________________
                         
    Training given by     _______________           _______________      
                                           Name                                 Sign/Date

    ANNEXURE II
    AIR VELOCITY MEASUREMENT TEST
    Date of Test         : __________________
    Equipment ID:
    Name of testing instrument Digital Anemometer
    Calibrated on    : ____________________        Calibration due: ___________

    HEPA
    FILTER No.
    Velocity [FPM]
    Average
    Velocity
    [FPM]
    Location 1    (V1)
    Location 2
    (V2)
    Location 3
    ( V3)
    Location 4
    (V4)
    Location 5
    (V5)
    1
    2
    3
    4
    5
    6

    Checked by
    (Production)   _________________   _________________    ____________
                                   (Name)                       (Signature)                      (Date)
    Verified by
    (QA)               _________________   _________________    ____________
                                   (Name)                       (Signature)                      (Date)


    ANNEXURE III
              HEPA FILTER INTEGRITY TESTING
    Date of Test  : ________________                                   
    Equipment ID :
    Name of Testing Instrument: Aerosol Photometer 
    Calibrated on : ________________ Calibration due: _______________

    HEPA Filter No.

    Upstream
    Concentration
    %age
    Downstream
    Concentration
    %age (NMT 0.01%)
    Remarks
    1
    100%
    2
    100%
    3
    100%
    4
    100%
    5
    100%
    6
    100%
    Checked by
    (Production)  _________________   _________________    ____________
                                   (Name)                       (Signature)                      (Date)
    Verified by
    (QA)              _________________   _________________    ____________
                                   (Name)                       (Signature)                      (Date)


    ANNEXURE IV
    AIR FLOW PATTERN TEST

    Date of Test                              :
    Equipment ID                             :
    Teat Conditions                          :  Static
    Name of the Testing Chemical    :   Titanium Tetra Chloride
    CONDITION
    CRITERIA
    YES/NO
    STATIC
    Whether the fumes of the titanium tetra chloride follow unidirectional path

    Checked by
    (Production)   _________________   _________________    ____________
                                   (Name)                       (Signature)                      (Date)
    Verified by
    (QA)              _________________   _________________    ____________
                                   (Name)                       (Signature)                      (Date)


    ANNEXURE V
    NON VIABLE PARTICLE COUNT TEST

    Date of Test              : ____________                         Equipment ID:
    HEPA Filter No           :
    Test Conditions         : Static                                                                     
    Name of Instrument  : Air Borne Particulate Counter
    Calibrated on            : ________________                                                          
    Calibration due          : ________________
    Location
       No.
    Particle
     Size
    Particulate Count
    Average
    (A=C1
    +C+--
    +Cn )/n
    Average
    In M3
    (A=A*
    35.2)
    Count 1
    (C1)
    Count 2
    (C2)
    Count3
    (C3)
    Count 4
    (C4)
    Count 5
    (C5)
    L 1
    ≥ 0.5µ
    ≥ 5.0µ
    L 2
    ≥ 0.5µ
    ≥ 5.0µ
    L 3
    ≥ 0.5µ
    ≥ 5.0µ
    L 4
    ≥ 0.5µ
    ≥ 5.0µ
    L 5
    ≥ 0.5µ
    ≥ 5.0µ

    ANNEXURE VI
    HEAT DISTRIBUTION STUDY
    TEST INSTRUMENT DETAILS TO BE USED FOR VALIDATION
    Test instrument name
    Model no
    Calibration done date
    Sensor type
    & quantity
    Make
    Calibration due date

    CYCLE NO
    Equipment name
    Date of experiment
    Equipment id no
    Equipment make
    Cycle start
    Cycle end
    Equipment location

    STERILIZATION CYCLE PARAMETERS









    Conveyor
    Start temperature
    Sterilization Hold time
    Conveyor
    Stop temperature
    Belt speed   in mm
    Sterilization temperature

                                  STERILIZATION CYCLE OBSERVATION

    TIME
    TEMPERATURE SENSORS
    1
    2
    3
    4
    5
    6
    7
    8
    9
    10
    11
    12
    13
    14
    15
    16
    MEAN
    Ple. Refer temp mapping date and print out of reading & temperature mapping data as attachment
       
     Inference __________________________________________________
    __________________________________________________
         

     Reviewed By                               Reviewed By                                  Approved By       Head Engineering                        Head Production                               Head -QA
                                                         
    ANNEXURE VII 
    HEAT PENETRATION STUDY
    TEST INSTRUMENT DETAILS TO BE USED FOR VALIDATION
    Test instrument name
    Model no
    Calibration done date
    Sensor type
    & quantity
    Make
    Calibration due date

    CYCLE NO
    Equipment name
    Date of experiment
    Equipment id no
    Equipment make
    Cycle start
    Cycle end
    Equipment location

    STERILIZATION CYCLE PARAMETERS
    Conveyor
    Start temperature
    Sterilization Hold time
          -----
    Conveyor
    Stop temperature
    Belt speed   in mm
          -----
    Sterilization temperature
            -----
          -----

    STERILIZATION CYCLE OBSERVATION
    TIME
    TEMPERATURE SENSORS
    1
    2
    3
    4
    5
    6
    7
    8
    9
    10
    11
    12
    13
    14
    15
    16
    MEAN
    Ple. Refer temp mapping date and print out of reading & temperature mapping data as attachment
         Inference_________________________________________________________________________________________________________________________________________________________
          Reviewed By                               Reviewed By                                  Approved By 
    Head Engineering                        Head Production                                   Head -QA

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App