    1.0 OBJECTIVE

    To check the operation of the equipment for reliable and accurate results.

    2.0 SCOPE

    This procedure is applicable to disintegration test apparatus, installed in the quality control department.

    3.0 RESPONSIBILITY

    3.1 Doing: Technical Assistant/Executive
    3.2 Checking: Executive /Assistant Manager

    4.0 ACCOUNTABILITY

    Head of the Department

    5.0 PROCEDURE

    Frequency: Once in a month
    5.1 Operate the instrument as per standard operating procedure.
    5.2 When indicator display 37°C, measure the temperature of the individual beaker with a thermometer and record the results in Annexure- I
    5.3 Also record the number of up down cycles per minute. Note down four reading for each bath and record the results in Annexure –I.
    5.4 Also measure the oscillation and record the results in Annexure –I.
    5.5 If the instrument is out of limit as in Annexure-I, follow the S.O.P.
    Note : Each instrument is calibrated with the respective serial number bath. Do not interchange the bath.

    6.0 ABBREVIATION

    6.1 °C = Degree centigrade
    6.2 no. = Number

    ANNEXURE-I
    QUALITY CONTROL LABORATORY
       OPERATIONAL CHECK REPORT OF DISINTEGRATION TEST APPARATUS
    Temperature setting : 37°C   Observed temp.    Beaker 1. _________ ( Limit : 35.0°C to 39.0°C)
                                                                            Beaker 2. _________ ( Limit : 35.0°C to 39.0°C)

    Sr. No
    Observed Up down cycles/minute
                       Remark
    01





    02

    03

    04

    Limit : 29 to 32 cycles/minute
    Measurement of  Oscillation  (Stroke length )
    Sr. No
                Oscillation
                       Remark

    Arm 1
    Arm 2




    01


    02


    03


    04


    Limit : 55 +/- 2 mm
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App