SOP for Calibration of High Performance Thin Layer Chromatography | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    SOP for Calibration of High Performance Thin Layer Chromatography

    Standard operating procedure to calibrate the High performance thin layer chromatography.

    1.0 OBJECTIVE

    To check the performance of High performance thin layer chromatography.

    2.0 SCOPE

    This procedure is applicable to calibrate high performance thin layer chromatography.

    3.0 RESPONSIBILITY

    3.1 Doing : Technical Assistant
    3.2 Checking : Executive /Manager

    4.0 ACCOUNTABILITY

    Head of the Department

    5.0 PROCEDURE

    5.1 Check the performance of instrument in terms of
    1. Linearity of spotting (Annexure-I)
    2. Reproducibility of spotting (Annexure-II)
    5.2 Prepare the mobile phase as per method given in Annexure-I.
    5.3 Prepare the solution as mentioned in Annexure-I
    5.4 Use 10 X 10 cm. HPTLC plate (Kieselgel 60F254 or Identical)
    5.5 For Linearity of spotting
    5.5.1 Apply 2 µl, 4µl, 6µl ,8µl and 10 µl of solution on HPTLC plate with spotter.
    5.5.2 Allow the plate to run in mobile phase.
    5.5.3 Dry the plate with drier.
    5.5.4 Scan the plate with scanner.
    5.5.5 Check the linearity and corelation coefficient.
    5.5.6 Fill in the Data as mentioned in Annexure-I
    5.6 Reproducibility of spotting
    5.6.1 Apply 10µl solution on HPTLC plate for five times in sequence
    5.6.2 Allow the plate to run in mobile phase.
    5.6.3 Dry the plate with drier.
    5.6.4 Scan the plate with scanner.
    5.6.5 Calculate the RSD for five track ( RSD Limit : N.M.T 3.0%)
    5.6.6 Fill in the data mentioned in Annexure-II


    6.0 ABBREVIATIONS

    6.1 %= Percentage
    6.2 RSD= Relative standard deviation
    6.3 N.M.T = Not more than
    6.4 N.L.T = Not less than
    6.5 µl = Microlitter
    6.6 HPTLC = High performance thin layer chromatography
    6.7 API= Active pharmaceuticals ingredient

    ANNEXURE-I
    Q. C. DEPARTMENT
    PERFORMANCE CHECK RECORD
    (LINEARITY OF SPOTTING)
    NAME OF THE INSTRUMENT : HIGH PERFORMANCE THIN LAYER CHROMATOGRAPHY
    MAKE                                            :
    MODEL NO                                  :
    IDENTIFICATION NO.               :
                Application Parameters :
                Plate Size                     : 10 X 10 cms.             Slit Width              :      
                Starting position           : 15 mm                        Migration distance :
                Band  width                 :  6 mm                         Saturation time      :
                Distance space             :  8 mm
                Sec/ ml                         :  15
                Inj.Volume                   :  2,4,6,8,10 ml
                Mobile Phase               : Methanol  : Ethyl acetate : Toluene : Ammonia
                                                              5     :          8          :      4      :   0.4
    Sample preparation : 20 mg. Famotidine -----> 100 ml ---> 3 ml----> 10 ml with glacial acetic acid. (working std.)
                                  Injection volume                     Area
                                           2.0
                                           4.0
                                           6.0
                                           8.0
                                         10.0   
         
    Linearity : ________  ( It should be linear)  Corelation Coefficient : ___ ( Not less than 0.9900)
                Remarks  :  Satisfactory/Not Satisfactory
                Performance check  done by           :                                                      Date  :    

                Checked by                                       :                                                     Date  :

                Next Performance check  due date :

    ANNEXURE-II
    Q. C. DEPARTMENT
    PERFORMANCE CHECK RECORD
    (REPRODUCIBILITY OF SPOTTING)
    NAME OF  THE INSTRUMENT : HIGH PERFORMANCE THIN  LAYER CHROMATOGRAPHY
    MAKE                                             :
    MODEL NO                                   :
    IDENTIFICATION NO.                :

                Track no.        Applied volume (µl)         Area
                   1                           10
                   2                           10 
                   3                           10
                   4                           10
                   5                           10

               RSD  =__________ ( Limit: Not more than 3.0%)
               Remarks                     : Satisfactory/Not Satisfactory
               Performance check  done by           :                                                    Date :    

               Checked by                                      :                                                    Date :

               Next Performance check  due date :
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App