    SOP for Calibration of Mercury Filled Glass Thermometer

    Standard operating procedure to calibrate the Mercury filled glass thermometer.

    1.0 OBJECTIVE

    To provide a standard operating procedure for the calibration of Mercury filled glass thermometer.

    2.0 SCOPE

    This procedure is applicable to all thermometer use in plant.

    3.0 RESPONSIBILITY

    3.1 Doing : Technical Assistant/Executive
    3.2 Checking : Executive /Manager

    4.0 ACCOUNTABILITY

    Head of the Department

    5.0 PROCEDURE

    Frequency : Once in a six month
    5.1 Assign a Q.C Report No to each thermometer as per following logic PL/0000 Where,
    PL= Abbreviation of Company and 0000 = serial number.
    5.2 Keep the thermometer which are to be calibrated at room temperature till room temperature is achieved.
    5.3 Visually inspect the thermometer for cracks, gaps, clear correct markings, chipped or distorted glass, and separation in the mercury column.
    5.4 If separation in the mercury (liquid) column, found then take following corrective action.
    5.4.1 Immerse the bulb portion of the thermometer in a block of dry ice until the meniscus has stopped moving. Ensure not to freeze the mercury (liquid) in the bulb.
    5.4.2 If the separation remains, gently tap the bulb on a soft object and swing the thermometer in a small arch until the mercury (liquid) is united.
    5.5 Thermometer with measurement range above 50°C are calibrated by third party.
    5.6 Calibration Procedure for thermometer having measuring range up to 50°C.
    5.6.1 First keep all 50°C range thermometer at room temperature till room temperature is achieved.
    5.6.2 After achieving room temperature, keep all thermometer with reference thermometer in a ice bath.
    5.6.3 Ensure the thermometer is immersed to the proper immersion depth.
    5.6.4 When meniscus has stopped moving gently tap the thermometer, as some amount of the meniscus may adhere to the wall of the thermometer.
    5.6.5 View the 0°C mark and compare test thermometer with reference thermometer.
    5.6.6 Record the value of test and reference thermometer in Annexure-I.
    5.6.7 Use low temperature water bath for calibration.
    5.7 For first temperature point use following procedure.
    5.7.1 Set the water bath at 10°C + 1 °C temperature. After achieving the temperature 10°C + 1 °C, keep reference calibrated thermometer and test thermometer inside the water bath.
    5.7.2 After 5 minutes observed the temperature of reference and test thermometer.
    5.7.3 Record the value of test and reference thermometer in Annexure-I.

    5.8 For second temperature point use following procedure.
    5.8.1 Set the water bath at 25°C + 1 °C. After achieving the temperature 25°C + 1 °C, keep reference calibrated thermometer and test thermometer inside the water bath.
    5.8.2 After 5 minutes observed the temperature of reference and test thermometer.
    5.8.3 Record the value of test and reference thermometer in Annexure-I.
    5.9 For third temperature point use following procedure.
    5.9.1 Set the water bath at 40°C + 1 °C. After achieving the temperature 40°C + 1 °C, keep reference calibrated thermometer and test thermometer inside the water bath.
    5.9.2 After 5 minutes observed the temperature of reference and test thermometer.
    5.9.3 Record the value of test and reference thermometer in Annexure-I.
    5.10 Ensure the results are within the tolerance limit.
    5.11 If values are out of tolerance, keep thermometer for 72 hrs. For stabilization and then recalibrate it. If it is out of limit, remove from the services and destroy the thermometer.
    5.12 Maintain the issue record of thermometer.

    Related: Different Types of Temperature Sensors

    6.0 ABBREVIATIONS

    6.1 SOP= Standard Operating Procedure
    6.2 °C= Degree centigrade


    ANNEXURE-I
    Q. C. DEPARTMENT
    THERMOMETER CALIBRATION RECORDS
    Calibrated Reference Thermometer No.  :____________________ of  0 oC to 50oC
    Ref. Thermometer Calibration date          :____________________
    Ref. Thermometer Certificate No.            :____________________
    Ref. Thermometer Calibration Due Date : ____________________
    TEMPERATURE OF STANDARD THERMOMETER AS PER CERTIFICATES:
      Temperature
     Temperature in  certificates
    0oC

    10oC

    25oC

    40oC



    RESULTS OBSERVED:
    Thermometer
    Q.C
    Report no.
    Thermometer Id.No.
    0oC
    10oC
    25oC
    40oC
    Remarks
    Tolerance
    Standard  thermometer
    +1.0 oC
    from
    temperature of
    reference thermometer
    Test Thermometer
    Calibration date:                                                                 Calibrated  by :

    Calibration due  date :                                                        Checked by :
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App