SOP for Operating Procedure of Fogster –ULV Fogger Machine | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    SOP for Operating Procedure of Fogster –ULV Fogger Machine

    Standard operating procedure Ultra Low Volume fogger machine to disinfectant the sterile area in aerosol state.

    1.0 OBJECTIVE

    To lay down the operating procedure of fogger machine.

    2.0 SCOPE

    This procedure is applicable to disinfectant the area in the aerosol state using fogger machine.

    3.0 RESPONSIBILITY

    3.1 Doing: Technical Assistant/ Executive
    3.2 Checking: Technical Assistant/Executive/Manager

    4.0 ACCOUNTABILITY

    Head of the Department
    FREQUENCY: Weekly

    5.0 PROCEDURE 

    5.1 Switch OFF the AHU of the sterility room & Inoculation room. Wear the mask, gloves & goggles before starting the work.
    5.2 Loosen tank clamp by pulling finger tabs upwards.
    5.3 Lift the power head of the tank & set it down on a clean surface.
    5.4 Add fogging liquid to tank (max up to 4 litres). Follow the instruction given in the table for the amount of disinfectant for specified area.
    5.5 Replace & reclamp the power head of the tank.
    5.6 Set the required flow control with the help of flow control value.
    Area
    ml of Disinfectant (super-25)
    Purified/distilled water
    Microbiology (Inoculation room)
    80 ml
    1600 ml
    Sterile area
    60 ml
    1200 ml
    Cooling zone
    30ml
    600 ml
    Airlock  III
    15 ml
    300 ml
    Airlock  II
    5 ml
    200 ml
    Airlock  I
    5 ml
    200 ml

    Calculation :- 

    For amount of disinfectant =L x W x h
    = For 1000 in cubic feet = 1 litre solution.
    Turn the value knob in counter clockwise direction for higher flow (heavy fog) & clockwise direction for droplets.
    5.7 Lock the value with rotating red nut for constant flow of liquid solution out of the fogger.
    5.8 Plug the fogger machine in mains.
    5.9 Switch on the mains.
    5.10 Place it at appropriate place in the room with nozzle direction pointed in desired position.
    5.11 Switch on the machine. So the machine generate the fog.
    5.12 After the completion of fogging treatment remove tank & add sterile distilled water into tank & operate for 2-3minutes to ensure cleaning of internal parts. Switch ‘OFF’ the machine & switch ‘OFF’ the mains.
    5.13 Clean the tank.

    Precautions :-

    1. Refer to material safety datasheets of chemical to be logged
    2. Ware mask, gloves & goggles.
    3. Donot use flammable liquid for fogging & never fog near or towards an open flame.
    4. Donot use corrosive liquid, paint lacquer & wettable powder solution.
    5. Donot allow fog, dirt or dust to enter the machine from the rear end.
    6. Always keep the tank clean.

    6.0 ABBREVIATIONS

    ULV = Ultra Low Volume
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App