SOP for Physical Monitoring of Microbiology Section | Pharmaguideline
  •   Pharmaguideline Courses Start Learning
  • Apr 17, 2021

    SOP for Physical Monitoring of Microbiology Section

    Standard operating procedure for monitor the temperature, differential pressure & UV light burning hours of micro section

    1.0 OBJECTIVE

    To monitor the temperature, differential pressure & UV light burning hrs of the micro section.

    2.0 SCOPE

    It is applicable to monitor and record the differential pressure, temperature & UV light burning in the various area of the micro section.

    3.0 RESPONSIBILITY

    3.1 Doing: Technical Assistant/ Executive
    3.3 Checking: Executive/Manager

    4.0 ACCOUNTABILITY

    Head of the Department

    5.0 PROCEDURE

    5.1 Before starting work observe the differential pressure of Magnahelic gauge of following room.

    Area
    Limit (Pascal)
    Airlock of Inoculation room
    NLT 15
    Inoculation room
    NLT 15
    Passage
    NLT  5
    Media preparation
    NLT 5
    Airlock I
    NLT 5
    Airlock II
    NLT 5
    Airlock III
    NLT 5
    Cooling zone
    NLT 15
    Sterility test room
    NLT 5
    5.2 Record the reading.
    5.3 “ON” the UV light of all area, after shift end & “off” the UV light before starting the work.
    5.4 Observe the UV light Burning hours from UV light Burning hours meter of the following area
    UV light Burning hours meter - box I
    1. Airlock II
    2. Airlock III
    3. Cooling zone
    On UV light Burning hours meter Box II
    1. Sterility Room
    2. Inoculation Room (Microbiology Room)
    3. Airlock I of Inoculation Room
    5.5 Record the observation.
    5.6 Replace the UV light after completion of 6000 hours of burning and record.
    5.7 Observe the temperature with calibrated thermometer of following area, three times a day.
    1. Walk-In –Incubator (30-35°C)
    2. Walk-In – Incubator ( 30-35°C)
    Record the result in Annexure-I.

    Acceptance criteria : -

    Walk-In-Incubator (30-35°C) --> Temperature should be 30-35°C
    Walk-In-Incubator (30-35°C) --> Temperature should be 30-35°C
    Also verify the temperature in circular chart recorder. Replace the chart with new one
    After 24 hours.

    6.0 ABBREVIATION

    UV light = Ultra Violet light
    NLT = Not less than

    ANNEXURE –I
    MICROBIOLOGY LAB, Q.C.
    TEMPERATURE RECORD OF WALK-IN INCUBATOR
    MONTH_________________


    DATE

    WALK IN INCUMBATOR
    (20 to 25°C)
    WALK IN INCUBATOR
    (30 to 35°C)
    SIGN
    &
    RRMARK
    Time à
    1
    3
    3
    1
    3
    3
    9:00
    13:00
    17:00
    9:00
    13:00
    17:00















































































           
               Acceptance criteria:
               Walk in-incubator    (20-25°C)  à  Temperature should be within 20-25°C
     Walk in-incubator (30-35°C)  à   Temperature should be within 30-35°C
    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories 🏷️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses 🎖️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS 📋

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App