Validation documents go beyond regulatory documentation; they are written proof that the pharmaceutical premises, equipment, utilities, computerized systems, analytical methods, and manufacturing methods work properly on a permanent basis. Validation document quality establishes, among other things, the trust level of regulatory authorities in the company’s performance and products.
However, many companies spend weeks on preparing validation protocols and reports but only a couple of hours on their review. Therefore, even such mistakes as incorrect acceptance criteria, poor traceability of information, calculation mistakes, unsubstantiated conclusions, and inconsistent data are usually revealed during inspection rather than in-house review.
In my opinion, efficient validation document review should not be considered as proofreading; it should rather be viewed as scientific evaluation of the document performance in terms of accomplishing the validation goal. In this case, the reviewer should scrutinize the assumptions made in the document, verify its supporting proof, and check whether the conclusions presented in the document are backed by the documented data.
This paper will consider some useful methods that validation experts may apply to make the validation document review successful before approval.
For instance:
"Results should be satisfactory."
Instead, try to write acceptance criteria in a measurable form. For instance:
The auditor should be able to trace:
User Requirement Specification (URS)
↓
Design Qualification (DQ)
↓
Installation Qualification (IQ)
↓
Operation Qualification (OQ)
↓
Performance Qualification (PQ)
↓
Running test
Similarly, each requirement should be linked to corresponding tests. Broken traceability often turns into a serious issue during inspection observations.
Some of the sources are:
1. Attributable
2. Legible
3. Contemporaneous
4. Original
5. Accurate
6. Complete
7. Consistent
8. Enduring
9. Available
Special attention must be paid to
1. Calculations
2. Electronic data
3. Audit trail
4. Consistency of time and dates
5. Versioning of documents
Weaknesses in data integrity can result in failure of validation studies.
"All tests were successful."
But write the claim:
- Purified Water
- PW
- P/W
- Water System
Pick one of the terms and stick to it.
A standardized checklist is also great for ensuring that vital details are not missed.
The act of reviewing validation papers is an integral part of pharmaceutical quality systems. The review can ascertain not just that the tests have been carried out but that the validation also has a scientific basis for the conclusion reached.
I have come to understand that the validation processes based on thorough document reviews result in a stronger system than the one based on repetitive validation studies. The improvement in the quality of validation documents can be attained only when reviewers evaluate not only acceptance criteria, raw data, and final conclusions but also whether the assumptions are correct, backed by evidence and the application of science in similar cases.
However, many companies spend weeks on preparing validation protocols and reports but only a couple of hours on their review. Therefore, even such mistakes as incorrect acceptance criteria, poor traceability of information, calculation mistakes, unsubstantiated conclusions, and inconsistent data are usually revealed during inspection rather than in-house review.
In my opinion, efficient validation document review should not be considered as proofreading; it should rather be viewed as scientific evaluation of the document performance in terms of accomplishing the validation goal. In this case, the reviewer should scrutinize the assumptions made in the document, verify its supporting proof, and check whether the conclusions presented in the document are backed by the documented data.
This paper will consider some useful methods that validation experts may apply to make the validation document review successful before approval.
Understand the Purpose Before Starting the Review
Before starting a protocol or report, make sure you understand what they intend to convey.For instance:
- Installation Qualification (IQ) indicates that the equipment is installed per approved specifications.
- Operational Qualification (OQ) shows that the equipment works within the specified range.
- Performance Qualification (PQ) acts as a proof of being operational under normal conditions.
- Process Validation shows that the process continuously produces products meeting the determinant quality characteristics.
- Cleaning Validation is a proof of the efficiency of cleaning processes in removing the residues of the product and cleaning agents.
Verify Compliance with the Validation Master Plan
All validation activities must be consistent with the validation master plan of the company. Make sure that this file follows:- Validation strategy approved
- Equipment classification
- Results of the risk assessment
- Requirements of the validation lifecycle
- Appropriate company practices
Review Document Completeness First
Prior to analyzing the technical part, make sure the document is complete. Make sure it includes:- Document number
- Revision number
- Title
- Scope
- Objective
- Roles
- References
- Approval signatures
- History of revisions
- Appendices
Check Technical Accuracy
Technical aspects need maximum focus. For example, the questions might be:- Is the goal of validation clear?
- Is the basis of acceptance criteria scientifically sound?
- Are the testing methods fit for the purpose?
- Are the sampling procedures appropriate?
- Are the calculations right?
- Is there evidence to support the conclusions reached?
Evaluate the Acceptance Criteria
Acceptance criteria must, at all times, be:- Quantifiable
- Scientifically based
- Regulatory compliant
- Product-specific if necessary
- Signed off prior to execution
"Results should be satisfactory."
Instead, try to write acceptance criteria in a measurable form. For instance:
- Temperature: 121 ± 2 degrees
- Relative humidity: 45-55%
- Differential pressure: 15 Pa or more
- Recovery: 70% at least
Ensure Traceability Throughout the Document
One of the important characteristics of a good validation document is full traceability.The auditor should be able to trace:
User Requirement Specification (URS)
↓
Design Qualification (DQ)
↓
Installation Qualification (IQ)
↓
Operation Qualification (OQ)
↓
Performance Qualification (PQ)
↓
Running test
Similarly, each requirement should be linked to corresponding tests. Broken traceability often turns into a serious issue during inspection observations.
Verify Supporting Data
Reviewers are supposed to check reported results against original records whenever it is possible.Some of the sources are:
- Calibration report
- Raw analytical data
- Instrumentation printout
- Environmental monitoring report
- Equipment log book
- Electronic audit trail
- Batch manufacturing records
Confirm Data Integrity
The validation documentation must indicate adherence to ALCOA principles. Accert the following data is:1. Attributable
2. Legible
3. Contemporaneous
4. Original
5. Accurate
6. Complete
7. Consistent
8. Enduring
9. Available
Special attention must be paid to
1. Calculations
2. Electronic data
3. Audit trail
4. Consistency of time and dates
5. Versioning of documents
Weaknesses in data integrity can result in failure of validation studies.
Review Deviations Carefully
Few validation studies do not have deviations. It is necessary that the reviewer finds out:- Has the deviation been documented?
- Was the investigation of the reasons for the deviation conducted?
- What corrective actions were taken?
- Did the deviation influence the results of the validation?
Assess Statistical Evaluation
Statistical evaluation forms an essential part of some validation reports. Consider the following:- Sample size
- Average value
- Standard deviation
- Capacity of a process
- Trend evaluation
- Confidence range
- Control charts
Evaluate Conclusions Critically
One of the most common problems with many validation reports is their conclusions. Try to avoid conclusions that only restate the results of the testing. Make sure, instead, that your conclusions help answer the initial question of the validation. For instance, avoid:"All tests were successful."
But write the claim:
- Whether acceptance criteria were satisfied
- Whether the system is adequate for the purpose of its usage
- Any limitations
- Recommendations on monitoring processes
- Whether qualification of the system is needed.
Review Language and Consistency
While grammatical correctness holds less weight than technical accuracy, proper communication makes any document better. Be sure to check for consistency in:- Identifiers
- Brands
- Numbers of batches
- Measurements
- Terms used
- Abbreviations used
- Purified Water
- PW
- P/W
- Water System
Pick one of the terms and stick to it.
Use a Structured Review Checklist
The structured review checklist guarantees that reviewers agree on critical points. Sample review items include:| Review Area | Questions |
| Scope | Is the objective clearly defined? |
| Protocol Compliance | Was every approved test executed? |
| Acceptance Criteria | Are limits scientifically justified? |
| Raw Data | Does source data support reported results? |
| Calculations | Have formulas been verified? |
| Deviations | Were all deviations assessed appropriately? |
| Conclusions | Are they supported by evidence? |
| Approvals | Are all required signatures present? |
A standardized checklist is also great for ensuring that vital details are not missed.
Common Review Deficiencies
Listed below are common errors that are found during inspection by various regulatory agencies:- Referring to protocols that have been copied without being checked.
- Using erroneous equipment identifiers.
- Not providing a sufficient amount of raw data.
- Presenting conclusions that have not been supported.
- Not having acceptance criteria in place.
- Not properly assessing deviations.
- Not being able to prove the results being tested.
- Not implementing proper revision control.
- All required approval signatures obtained before completion of the required review process.
Strengthen the Review Process
To improve the quality of validation documents, organizations can do the following things:- Assigning technically competent reviewers.
- Making sure that technical review is kept apart from format review.
- Utilizing independent reviewers when available.
- Conducting peer reviews prior to the QA.
- Looking at past validation results.
- Training reviewers about risk-based approach.
- Keeping the validation templates uniform.
- Conducting audits of validation documents from time to time.
The act of reviewing validation papers is an integral part of pharmaceutical quality systems. The review can ascertain not just that the tests have been carried out but that the validation also has a scientific basis for the conclusion reached.
I have come to understand that the validation processes based on thorough document reviews result in a stronger system than the one based on repetitive validation studies. The improvement in the quality of validation documents can be attained only when reviewers evaluate not only acceptance criteria, raw data, and final conclusions but also whether the assumptions are correct, backed by evidence and the application of science in similar cases.