SOP for Calibration of Thermometer in Pharma

Standard operating procedure to calibrate the thermometer using reference standards having standard melting range.

1.0 OBJECTIVE

To lay down the procedure for calibration of Thermometer.

2.0 SCOPE

This SOP shall be applicable to Quality Control Dept.

3.0 RESPONSIBILITY

Quality Control Supervisor

4.0 ACCOUNTABILITY

Sr. Manager Quality Assurance

5.0 PROCEDURE

5.1 Each thermometer shall be numbered and identified by its number whenever it is used, either for calibration or for routine temperature measurements. These thermometers are calibrated using following reference standards for their accuracy and record:
Sr. No.
Substance
Standard Thermometer
Reading
Standard Melting Range
1.
Vanillin

81 – 83°C
2.
Acetanilide

114 – 116°C
3.
Sulphanilamide

164.5 –166.5°C
4.
Caffeine

234 – 237°C
Any thermometer that breaks or does not respond to given temperature range should be discarded.

6.0 ABBREVIATION

6.1 SOP: Standard Operating Procedure

Get ready to use editable documents in MS-Word FormatView List





Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of Pharmaceutical Guidelines, a widely-read pharmaceutical blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
.moc.enilediugamrahp@ofni :liamENeed Help: Ask Question


Click Here

No comments: Read Comment Policy ▼

Post a Comment


Follow Pharmaguideline


SPONSORED POST

SPONSORED POSTS

CURRENT JOBS

Show All ❭❭Jobs by PharmaJobs

WRITE A POWERFUL CV

Click Here



ADVERTISE HERE


GET APP FOR NEWS UPDATES

Scan to Download

Android App
Android App

Recent Posts