    SOP for Cleaning of Dissolution Tester

    Standard operating procedure to clean the dissolution tester including bowls, baskets and paddles.

    1.0 OBJECTIVE

    To lay down the procedure for cleaning.

    2.0 SCOPE

    This procedure is applicable to dissolution tester installed in Quality Control Laboratory.

    3.0 RESPONSIBILITY

    3.1 Doing: Worker
    3.2 Checking: Executive/ Manager

    4.0 ACCOUNTABILITY

    Head of the Department

    5.0 PROCEDURE

    5.1 Lift the Stirrer Unit by UP(­) key.
    5.2 Switch ‘OFF’ the heater.
    5.3 Switch ‘OFF’ the instrument.
    5.4 Switch ‘OFF’ the main.
    5.5 Cleaning procedure for Apparatus
    Frequency: After completion of every dissolution test.
    5.6 Remove the bowl from its place and discard the medium of each bowl. Wash each bowl properly with R.O. water then keep it inside the dissolution bath number wise.
    5.7 Remove the basket/paddle apparatus from the dissolution tester. Wash them with R.O. water and dry with tissue paper and keep at the proper place.
    5.8 Keep the record in the instrument log card.
    5.9 Cleaning procedure for water bath of Dissolution Tester.
    Frequency: Once in a week or if necessary.
    5.10 Remove all the bowls from the water bath.
    5.11 Remove the baskets/paddle apparatus from the stirrer unit.
    5.12 Remove the circulating pipe which is provided at the right rear side of the water bath (inlet)
    5.13 Start the Circulating Pump from the options menu and drain the dirty water in wastage water bucket.
    5.14 Insert the Circulating Pipe in its proper place.
    5.15 Wash the bath with raw water, detergent solution and with raw water sequentially, until no foam of detergent remains in the bath. After complete removal of the detergent, wash the bath finally with R.O. water
    5.16 Fill the bath with R.O. water up to the level mark.
    5.17 Record it as per Annexure-I.

    Related: Calibration of Dissolution Tester

    6.0 ABBREVIATIONS

    6.1 SOP - Standard Operating Procedure
    6.2 R.O. water - Reverse Osmosis Water.


    ANNEXURE-I
    Q.C. DEPARTMENT
    CLEANING PROCEDURE FOR DISSOLUTION TEST APPARATUS
    NAME OF THE INSTRUMENT :
    MAKE                                           :
    MODEL NO                                  : TDT-08  L
    IDENTIFICATION  NO.

    Date  of cleaning
    Due date of cleaning
    Cleaned by
    Checked by
    Remarks








































    Get editable documentsView List

    No comments:

    Post a Comment

    Please don't spam. Comments having links would not be published.


    Popular Categories ๐Ÿท️

    QA SOPs QC SOPs Micro SOPs HVAC Production SOPs Stores SOPs Checklists Maintenance SOPs HPLC Sterile GLP Validation Protocols Water System GDP Regulatory Maintenance Calibration Warning Letters Education B.Pharmacy Regulatory Updates
    Online Courses Salary Calculator
    Follow on Whatsapp


    Certificate Courses ๐ŸŽ–️

    Pharmaceutical Courses

    Follow Pharmaguideline


    DOCUMENTS ๐Ÿ“‹

    PHARMACEUTICAL DOCUMENTS




    Editable Pharmaceutical Documents in MS-Word Format. Ready to use SOPs, Protocols, Master Plans, Manuals and more...

    View


    adsbypg


    GET APP FOR NEWS UPDATES


    Android App