1.0 OBJECTIVE
To
calibrate the equipment for reliable and accurate results.
2.0 SCOPE
This procedure is applicable to calibrate bursting
strength tester.
3.0 RESPONSIBILITY
3.1 Doing
: Technical Assistant/ Executive
3.2 Checking
: Executive /Manager
4.0 ACCOUNTABILITY
Head
of the Department
5.0 PROCEDURE
Frequency
of calibration : Once in three month
5.1 Operate
the instrument as per SOP
5.2 Place the standard piece of foil
over the diaphragm and take the reading from gauge.
5.3 The result should within the standard foil
range. If any results differ by more than specified range
limit. Follow the S.O.P.
6.0 ABBREVIATIONS
6.1 kg
= Kilogram
6.2 cm.
= Centimeter
6.3 S.O.P.= Standard operating procedure
6.4 No.= Number
ANNEXURE-I
CALIBRATION REPORT OF BURSTING STRENGTH TESTER
CALIBRATION DATE
|
DATE OF LAST CALIBRATION DONE
|
NEXT DUE FOR
CALIBRATION
|
INSTRUMENT DETAILS
| ||
INSTRUMENT NAME
| INSTRUMENT MAKE |
INSTRUMENT
IDENTIFICATION NO.
|
OBSERVATION
|
Certificate No. of Std. Foil used : (1) (2)
Sr. No.
|
Specified value Range of Std. Foil
|
Observed value
|
01
| ||
02
|
Acceptance Criteria : Observed value should be with in the specified range of standard foil value.
Remarks
Calibration date
Calibration due date
Calibrated by
Checked by
|
:
:
:
:
:
|