SOP for Calibration of Infrared Moisture Balance

Standard operating procedure to calibrate the infrared moisture balance comparing with LOD at 105 deg.C.

1.0 OBJECTIVE

To calibrate the equipment for reliable and accurate results.

2.0 SCOPE

This procedure is applicable to infrared moisture balance, installed in the quality control department.

3.0 RESPONSIBILITY

3.1 Doing: Technical Assistant
3.2 Checking: Executive /Manager

4.0 ACCOUNTABILITY

Head of the Department

5.0 PROCEDURE

Frequency: Once in a three month
5.1 Operate the instrument as per SOP.
5.2 Set the temperature by ‘voltage regulator knob’.
5.3 After fifteen minute, observe the temperature and record it in the annexure.
5.4 Select any raw material like starch or any in-process granules sample. Carry out LOD at 105oC. Calculate and record % LOD in annexure-I
5.5 For the same sample perform its LOD in the infrared moisture balance at the same temperature and record the result in Annexure.
5.6 The performance of the instrument is satisfactory if the observed % L.O.D. difference between the oven and I.R.balance is within the tolerance limit of 10 % of the % LOD observed in the oven, otherwise follow the S.O.P.

6.0 ABBREVIATIONS

6.1 % = Percentage
6.2 LOD = Loss on drying
6.3 I.R. = Infrared
Get ready to use editable SOPs in MS-Word FormatView List





is a prominent Pharmaceutical Quality Assurance expert, consultant and the founder of Pharmaguideline. With over 22 years of hands-on experience in cGMP-compliant manufacturing environments, he specializes in establishing validation protocols, sterile area controls and data integrity systems. Ankur routinely interprets international regulatory frameworks (including FDA, EMA and ICH guidelines) to help global pharmaceutical professionals ensure strict regulatory compliance and operational excellence. Connect with Ankur on LinkedIn. Need Help: Ask Question