SOP for Calibration of pH Meter Temperature Sensor

Standard operating procedure for the calibration of pH meter temperature sensor using water for injection (WFI) / purified water (PW).

1.0 Objective

To provide a standard operating procedure for the calibration of pH meter temperature sensor.

2.0 Scope

This procedure is applicable to pH meter.

3.0 Responsibility

3.1 Doing: Technical Assistant/Executive
3.2 Checking: Executive /Manager

4.0 Accountability

Head of the Department

5.0 Procedure

Frequency: Twice in Year.
5.1 Calibration Procedure for pH meter temperature sensor
5.1.1 Keep the sensor probe of reference calibrated digital thermometer in the beaker containing water for injection (WFI) / purified water (PW) of about temperature 35°C.
5.1.2 Add cold WFI/PW slowly drop wise and bring the temp. between 27°C to 30°C.
5.1.3 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.
5.1.4 Add WFI/PW in some more quantity slowly drop wise to bring the temp. between 26°C to 23°C.
5.1.5 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.
5.1.6 Add cold WFI/PW slowly drop wise and bring the temp. between 20°C to 22°C.
5.1.7 Compare the temp. shown by reference calibrated digital thermometer and pH meter & record it in format as Annx. I.
5.1.8 Acceptance Criteria : ± 3°C

6.0 Abbreviations

6.1 °C= Degree centigrade

ANNEXURE-I
Q. C. Department
PH METER TEMPERATURE SENSOR RECORDS
                                                                   
REF.No. :_________________
ISSUE DATE: _________________
ISSUED BY :____________________   



Calibrated  Ref. Digital Thermometer No.:____________________ 
Ref. Thermometer Calibration date          :____________________
Ref. Thermometer Certificate No.            :____________________
Ref. Thermometer Calibration Due Date  : ____________________
pH meter ID.No.                                      : ___________________
Results Observed:
Set Range
Observation
Difference
tolerance
Temperature as read on pH meter
Temperature as on read on display of Ref. Thermo.
30°C to 27°C



±0.3°C
26°C to 23°C



22°C to 20°C




Remarks                 :
Calibration date          :
Next Calibration date  :
Calibrated  by             :
Checked by                 :
Get ready to use editable SOPs in MS-Word FormatView List





is a prominent Pharmaceutical Quality Assurance expert, consultant and the founder of Pharmaguideline. With over 22 years of hands-on experience in cGMP-compliant manufacturing environments, he specializes in establishing validation protocols, sterile area controls and data integrity systems. Ankur routinely interprets international regulatory frameworks (including FDA, EMA and ICH guidelines) to help global pharmaceutical professionals ensure strict regulatory compliance and operational excellence. Connect with Ankur on LinkedIn. Need Help: Ask Question