Testing of Water Samples
• pH: -
1. Use pH meter previously calibrated using three point calibration method at pH
2. 4.0, pH 7.0 and pH 10.0 using NIST traceable standard buffers.
3. Calibration Frequency: Daily
4. Acceptance Criteria: Variance NMT 0.05
5. Always preserve the electrode in saturated Kcl solution.
6. Rinse clean the electrode with purified water
7. While testing ensure tem…
2. 4.0, pH 7.0 and pH 10.0 using NIST traceable standard buffers.
3. Calibration Frequency: Daily
4. Acceptance Criteria: Variance NMT 0.05
5. Always preserve the electrode in saturated Kcl solution.
6. Rinse clean the electrode with purified water
7. While testing ensure tem…