CALIBRATION OF TOP PAN BALANCE
Calibration procedure
Internal calibration
It shall be performed as per the manufacturer’s instructions.Drift check
For the calculation of the drift 10 mg weight shall be used. Note down the 10 measurements of the 10 mg weight in the performance check log. (Attachment no. 01) Variation in the observed weight from the mean value does not exceed ± 0.2 mg.
 Performance check
Performance check
After the Auto calibration put 1 mg, 2 mg, 5 mg, 10 mg and 20 mg weights individually. The measurement shall be within the 0.1% of the actual mass value of the individual weight as given in the performance check log. (Attachment no. 02)
Put the external weight of 10 mg on the pan and note the 10 measurements. (Attachment no. 03)
Calculate the measurement of uncertainty as follows.
Measurement Uncertainity = Standard Deviation X 2
Actual Mass Value
Measurement of uncertainty shall be Not More Than 0.10.
Measurement uncertainty check
The measurement of uncertainty shall be carried out by using 10 mg weight.Put the external weight of 10 mg on the pan and note the 10 measurements. (Attachment no. 03)
Calculate the measurement of uncertainty as follows.
Measurement Uncertainity = Standard Deviation X 2
Actual Mass Value
Measurement of uncertainty shall be Not More Than 0.10.
Calibration frequency
It shall be performed daily, after any maintenance, after the relocation and after the power failure of the balance.
Each manufacturer supplies the maximum loading condition of the balance. The calibration of the balance shall be designed in such a way that the performance check weights cover the entire loading range of the balance. Lower and higher load limits shall be checked for the performance check. Before calibration make sure that the level bubble is in the center of the indicator and if internal calibration facility is available in the balance then it shall also be performed daily before performance check of the balance.
Always use a check weight appropriate for the particular balance. E.g. a 100 – mg check weight might be selected for a balance that has a load limit of 150 mg. Thus the operator must know the maximum capacity of the balance to select the correct check weight.
Attachment-1: Performance Check Log of Microbalance
        
          
Always use a check weight appropriate for the particular balance. E.g. a 100 – mg check weight might be selected for a balance that has a load limit of 150 mg. Thus the operator must know the maximum capacity of the balance to select the correct check weight.
Attachment-1: Performance Check Log of Microbalance
Instrument:_____________________                                  Code No :______________
Make         :_____________________                                 Model   :_______________  
| 
Date | 
Level indicator check | 
1 mg | 
2 mg | 
5 mg | 
10 mg | 
20 mg | 
Weight box no. and valid upto | 
Done by | 
Checked 
by | 
Remark | |
| 
Std. Wt. Cal No. Þ | |||||||||||
| 
Limit 
Þ | 
Bubble should be in Center | 
+ 0.1% of 
 Std. Weight 
(        to        ) | 
+ 0.1% of 
 Std. Weight 
(        to        ) | 
+ 0.1% of  
Std. Weight 
(        to        ) | 
+ 0.1% of Std. Weight 
(        to        ) | 
+ 0.1% of Std. Weight 
(        to        ) | |||||
Attachment-2: Auto Performance Check Log of Microbalance
Instrument:_____________________                             Code No :______________
Make         :_____________________                             Model   :_______________  
Check level indicator (Bubble should be in center) : _________________________________
                                                                  (Paste Performance check chart here)
Calibrated by : _______________                                               Date : ____________________
Checked  by : _______________                                                Date : ____________________
Attachment-3: Measurement Uncertainity Check of Microbalance
Instrument:_____________________                   Code No :______________
Make         :_____________________                   Model   :_______________  
| 
Date | 
Std. Wt. | 
Std. Wt. Cal. No. | 
Observation | 
 of Uncertainty  
Less than or equal to 0.001 | 
Weight box no. and valid upto | 
Done by | 
Checked  by | 
Remark | ||||
| 
10 mg | ||||||||||||
Calculation 
Measurement = Standard Deviation X 2
Measurement = Standard Deviation X 2
uncertainty               Actual Mass Value


What do u mean by actual mass value in measurement of uncertainty?
ReplyDeleteIt means actual weight.
DeleteExactly, it is certified weight of the standard being used.
Deletehello sir,
ReplyDeleteverification of balance is necessary after every power failuar.
Reply plz. From syed sabir
Yes and if you change the location or move the balance you need to do calibration
Delete