Calibration of Gas Chromatography (GC)


Learn how to calibrate the Gas Chromatography (Carrier Gas Flow, Column oven temperature and Detector calibration).

Calibration of Carrier Gas Flow setting-

Touch on “tool” at right hand corner of screen on instrument by using stick provided with instrument.
Select option utilities.
Touch on column flow calc.
Enter column length, diameter, flow and column oven temperature; it will display required pressure to set by using carrier pressure knob.

Column oven temperature calibration:

Ware goggle and gloves while handling glass wool.
Insert the calibrated digital thermometer from topside by piercing glass wool.
Set column oven temperature on GC to 50°C, 100°C, 200°C, and 250°C.
Allow to stabilize the set temperature for 5 min and record temperature on digital thermometer at respective temperatures.
Frequency - Once in three months.

Detector Calibration:

Calibration of detector
Attach SS packed column of length 2 M x diameter 1/8 inch with stationary phase 
10 % C-20 M (Ch-W/HP) 80/100
Prepare 1.0% v/v toluene solution in n-Hexane
Set the chromatographic conditions as follows
Injector temperature  : 180°C
Detector temperature  : 220°C
Column flow  : 2.7 ml/min
Oven temperature  : 110°C (Isothermal)
Run Time  : 5 minutes
Inject blank and then six injections of 1.0% toluene solution in system and record chromatogram for all injections
Calculate the RSD for the area of six 1.0 % toluene injections.
The RSD should not be more than 15.0%
Frequency - Once in six months.

SET TEMPERATURE ON GC
OBSERVED TEMPERATURE ON DIGITAL THERMOMETER
ACCEPTANCE CRITERIA
50°c
        48-52°C
100°c
        98-102°C
200°c
        198-202°C
250°c
        248-252°C